Year : 1991  |  Volume : 39  |  Issue : 2  |  Page : 50-52

Prediction of intra ocular lens power in 100 patients using the ophthasonic a scan unit


Department of Ophthalmology, Baroda Medical College, Gujarat, India

Correspondence Address:
N K Limdi
Department of Ophthalmology, Baroda Medical College, Gujarat
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


PMID: 1916979

Rights and PermissionsRights and Permissions

The aim of this study was to determine the range of ametropias encountered by using the computerized Ophthasonic A-Scan. The SRK formula was used to compute the IOL power using axial length and keratometry as inputs. The accuracy of prediction of IOL power was calculated by noting the disparity between the expected post-operative refraction and the actual refraction obtained after one and a half months of IOL implantation. The results obtained were encouraging; the maximum deviation from the expected result being +/- 3.2 D.