Intraoperative cracking of the AMO Phacoflex-II Silicone (SI-40NB) intraocular lens while implanting with the Unfolder TM silver series system using hydroxypropylmethylcellulose


Dr. Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences, New Delhi, India

Correspondence Address:
Ankur Sinha
Dr. Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences, New Delhi
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0301-4738.33836

Rights and Permissions

How to cite this article:
Saxena R, Sinha A, Khokhar S, Dhawan M, Turuka K. Intraoperative cracking of the AMO Phacoflex-II Silicone (SI-40NB) intraocular lens while implanting with the Unfolder TM silver series system using hydroxypropylmethylcellulose. Indian J Ophthalmol 2007;55:396-7

How to cite this URL:
Saxena R, Sinha A, Khokhar S, Dhawan M, Turuka K. Intraoperative cracking of the AMO Phacoflex-II Silicone (SI-40NB) intraocular lens while implanting with the Unfolder TM silver series system using hydroxypropylmethylcellulose. Indian J Ophthalmol [serial online] 2007 [cited 2018 Dec 15];55:396-7. Available from: http://www.ijo.in/text.asp?2007/55/5/396/33836

Dear Editor,

Silicone and acrylic intraocular lenses (IOLs) are commonly used after phacoemulsification. Damage to acrylic [1],[2] and Clariflex [3] IOLs during implantation has been reported; most of them do not affect visual outcome but may need explantation. [2] We report cracking of Phacoflex-II during insertion, using hydroxylpropylemethyl cellulose (HPMC) with Unfolder TM system. [4]

A 56-year-old male presented with bilateral immature senile cataract with best corrected visual acuity (BCVA) of 20/80 in right eye and 20/40 in left eye. Detailed evaluation was unremarkable. Phacoemulsification was performed in the right eye.

Phacoflex-II was loaded into Unfolder TM using HPMC according to the manufacturer's instructions. Leading haptic was implanted successfully, while crack was noted at the trailing haptic-optic junction. The IOL optic had opened and it was impossible to withdraw. In spite of the crack, the IOL was well centered in the bag [Figure - 1]. On the first post surgical day, the patient had BCVA of 20/20. Crack measured approx. 2.4 mm and was not in the visual axis.

After six weeks, the IOL remained stable and well centered, without compromising BCVA.

There are reports [1],[2],[3] of IOL damage during implantation. Among silicone IOLs this complication has only been reported once with Clariflex but has never been reported with Phacoflex-II.

We have been using HPMC for insertion of Phacoflex-II using Unfolder TM for the past three years without any damage to the IOL. Though manufacturers instruct to avoid use of Healon TM with Unfolder TM , its use without any lens damage has been reported. [5]

We used HPMC for cost constraint. This was the first complication of IOL damage.

In our case, proper technique for injection was followed, with use of HPMC. Hydroxylpropylemethyl cellulose has lower pseudo-plasticity [6] than other viscoelastics, hence it is more difficult to begin and maintain injection through the cannula. [6] With HPMC, the force required to dislodge optic and haptic from the cartridge at initiation of injection is higher. This may lead to movement of optic followed by movement of trailing haptic after a time lag, due to inherent design of the injector. These factors could induce stress at the optic-haptic-junction leading to the crack.

These small peripheral cracks have no ill-effects on the visual outcome and stability of the IOL in the short term. If HPMC is used, a standby IOL should be ready at the time of implantation.

 
  References Top

1.
Balasubramanya R, Rani A, Dada T. Forceps-induced cracking of a single-piece acrylic foldable intraocular lens. Ophthalmic Surg Lasers Imaging 2003;34:306-7.  Back to cited text no. 1
[PUBMED]    
2.
Lee GA, Dal Pra ML. Cracked acrylic intraocular lens requiring explantation. Aust N Z J Ophthalmol 1997;25:71-3.  Back to cited text no. 2
[PUBMED]    
3.
Chuah JL, Rajesh CV. Intraoperative fracture of AMO Clariflex siliconee posterior chamber IOL. J Cataract Refract Surg 2005;31:1260-1.  Back to cited text no. 3
[PUBMED]  [FULLTEXT]  
4.
Olson R, Cameron R, Hovis T, Hunkeler J, Lindstrom R, Steinert R. Clinical evaluation of the Unfolder. J Cataract Refract Surg 1997;23:1384-9.  Back to cited text no. 4
[PUBMED]    
5.
Singh AD, Fang T, Rath R. Cartridge cracks during foldable intraocular lens insertion. J Cataract Refract Surg 1998;24:1220-2.  Back to cited text no. 5
[PUBMED]    
6.
Liesegang TJ. Viscoelastic substances in ophthalmology. Surv Ophthalmol 1990;34:268-93.  Back to cited text no. 6
[PUBMED]    


    Figures

  [Figure - 1]



 

Top
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
References
Article Figures

 Article Access Statistics
    Viewed2275    
    Printed19    
    Emailed0    
    PDF Downloaded141    
    Comments [Add]    

Recommend this journal