ORIGINAL ARTICLE
Year : 2010  |  Volume : 58  |  Issue : 3  |  Page : 219-222

Comparison of Eyemetrics and Orbscan automated method to determine horizontal corneal diameter


Vasan Eye Care Hospital, Chennai, India

Correspondence Address:
Arvind Venkataraman
Vasan Eye Care Hospital, M - 77 III Avenue, Anna Nagar East, Chennai - 600 102
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0301-4738.62647

Rights and Permissions

Purpose: To compare horizontal corneal diameter measurements using the Orbscan Eyemetrics function and Orbscan corneal topographer. Materials and Methods: Seventy-three eyes of 37 patients were included in the study. In all cases, the automated white-to-white (WTW) measurements were obtained using Orbscan by two observers. Using the Eyemetrics function, the WTW was measured manually by the same observers from limbus to limbus using the digital caliper passing through the five point corneal reflections on the Orbscan real image. The data was analyzed using SPSS software for correlation, reliability and inter-rater repeatability. Results: The mean horizontal corneal diameter was 11.740.32mm (SD) with the Orbscan and 11.920.33mm (SD) with Eyemetrics Software-based measurement. A good positive correlation (Spearman r = 0.720, P = 0.026) was found between these two measurements. The coefficient of inter-rater repeatability was 0.89 for the Orbscan and 0.94 for the Eyemetrics software measurements on the anterior segment images. The Bland and Altman analysis showed large limits of agreement between Orbscan WTW and Eyemetrics WTW measurements. The intra-session repeatability scores for repeat measurements for the Orbscan WTW and Eyemetrics measurements were good. Conclusion: Eyemetrics can be used to measure WTW and the Eyemetrics measured WTW was longer than the WTW measured by Orbscan.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed3020    
    Printed59    
    Emailed0    
    PDF Downloaded215    
    Comments [Add]    
    Cited by others 6    

Recommend this journal