LETTER TO THE EDITOR
Year : 2013  |  Volume : 61  |  Issue : 2  |  Page : 82-83

Descemet's tear due to injector cartridge tip deformity: Cartridge shaft deformity


Department of Ophthalmology, Vasantrao Naik Government Medical College, Yavatmal, Maharastra, India

Date of Web Publication15-Feb-2013

Correspondence Address:
Rajesh S Joshi
77, Panchatara Housing Society, Manish Nagar, Somalwada, Nagpur - 440 015, Maharastra
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0301-4738.107205

Rights and Permissions

How to cite this article:
Joshi RS. Descemet's tear due to injector cartridge tip deformity: Cartridge shaft deformity. Indian J Ophthalmol 2013;61:82-3

How to cite this URL:
Joshi RS. Descemet's tear due to injector cartridge tip deformity: Cartridge shaft deformity. Indian J Ophthalmol [serial online] 2013 [cited 2024 Mar 28];61:82-3. Available from: https://journals.lww.com/ijo/pages/default.aspx/text.asp?2013/61/2/82/107205

Dear Editor,

I read with interest an article by Biswas et al., on descemet's tear due to injector cartridge deformity. [1] I congratulate the authors for the suggestions given to avoid such a mishap. The authors described the situation were injector used was probably reusable for hydrophobic intraocular lenses (IOL).

I would like to share a situation were the cartridge shaft broke apart during the implantation of the hydrophilic IOL. These IOLs are provided with the disposable injector and the cartridge packed along with the IOL or separately. The leading haptic and optic went inside the capsular bag but the cartridge shaft broke away vertically parallel to the axis of the cartridge 6-8mm from the tip as the trailing haptic passed through the weak area [Figure 1] and [Figure 2]. The folding of the IOL was done by the author himself under the operating microscope. Another injector and the cartridge were used for the implantation of the IOL.
Figure 1: Deformed cartridge with protruding trailing haptic of the IOL (Arrow)

Click here to view
Figure 2: Weak area in the shaft of the cartridge (Arrow)

Click here to view


The mishap did not affect the postoperative visual out come, as the IOL was withdrawn from the anterior chamber but left us in doubt about the manufacturing defect. The matter was reported to the concerned manufacturer and requested to provide an extra injector and the cartridge.

I do agree with the authors about inspection of cartridge under the slit lamp/microscope for the damage so as to replace immediately. But weak areas on the cartridge may be missed sometimes. Such occurrences could be routine but few are reported. India being a huge market for the disposables used for cataract surgery quality control is essential. I take this opportunity to request All India Ophthalmological Society to form a quality control committee which will ensure uniform quality of the disposables used in the ocular surgery.

 
  References Top

1.
Biswas P, Sengupta S, Paul A, Kochgaway L, Biswas S. Descemet's tear due to injector cartridge tip deformity. Indian J Ophthalmol 2012;60:218-20.  Back to cited text no. 1
[PUBMED]  Medknow Journal  


    Figures

  [Figure 1], [Figure 2]


This article has been cited by
1 Comparative evaluation of an automated preloaded delivery system with a non-preloaded system
RajeshS Joshi
Indian Journal of Ophthalmology. 2022; 70(12): 4307
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
References
Article Figures

 Article Access Statistics
    Viewed2231    
    Printed54    
    Emailed0    
    PDF Downloaded171    
    Comments [Add]    
    Cited by others 1    

Recommend this journal