ORIGINAL ARTICLE
Year : 2018  |  Volume : 66  |  Issue : 3  |  Page : 402-406

Comparison of anterior segment measurements using Sirius Topographer® and Nidek Axial Length-Scan® with assessing repeatability in patients with cataracts


1 Department of Ophthalmology, School of Medicine, Afyon Kocatepe University, Afyonkarahisar, Turkey
2 Department of Ophthalmology, Manisa State Hospital, Manisa, Turkey

Correspondence Address:
Dr. Resat Duman
Department of Ophthalmology, School of Medicine, Afyon Kocatepe University, Afyonkarahisar
Turkey
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijo.IJO_859_17

Rights and Permissions

Purpose: The purpose of this study is to evaluate anterior segment measurements obtained using CSO Sirius Topographer® (CSO, Firenze, Italy) and Nidek Axial Length (AL)-Scan® (Nidek CO., Gamagori, Japan). Methods: A total of 43 eyes of 43 patients were included in this prospective study. The central corneal thickness (CCT), anterior chamber depth (ACD), white-to-white distance (WTW), flat keratometry (K1), steep keratometry (K2), and mean keratometry (K) values were randomly measured three times with each device by the same examiner. The intraclass correlation coefficient of repeatability was analyzed. The compatibility of both devices was evaluated using the 95% limits of the agreement proposed by Bland and Altman. Results: Examiner achieved high repeatability for all parameters on each device except the WTW measured by Sirius. All measurements except WTW and K1 taken with the Sirius were higher than that taken with the Nidek AL-Scan®. The difference in CCT, ACD, and WTW values was statistically significant. Conclusion: High repeatability of the measurements was achieved on both devices. Although Km, K1, and K2 measurements of the Sirius and the AL-Scan® showed good agreement, WTW, CCT, and ACD measurements significantly differed between two devices. Thus, anterior segment measurements except for Km, K1, and K2 cannot be used interchangeably between Sirius and Nidek AL-Scan® devices.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed2020    
    Printed20    
    Emailed0    
    PDF Downloaded234    
    Comments [Add]    
    Cited by others 6    

Recommend this journal