ORIGINAL ARTICLE
Year : 2018  |  Volume : 66  |  Issue : 9  |  Page : 1262-1267

Repeatability of ARK-30 in a pediatric population


1 Optometry Research Group, IOBA Eye Institute, School of Optometry, University of Valladolid, Valladolid, Spain; Vision Rehabilitation Lab, Department of Physics and Optometry, University of Minho, Braga, Portugal
2 Universidad de Valladolid, Instituto Universitario de Oftalmobiología Aplicada (IOBA); Hospital Clínico Universitario de Valladolid, Valladolid, Spain
3 Optometry Research Group, IOBA Eye Institute, School of Optometry, University of Valladolid; Departamento de Física Teórica, Atómica y Óptica, Universidad de Valladolid, Valladolid, Spain; School of Health Professions, Plymouth University, Derriford Road, Plymouth, UK

Date of Submission24-Mar-2018
Date of Acceptance31-May-2018
Date of Web Publication20-Aug-2018

Correspondence Address:
Dr. Raul Martin
School of Health Professions, Faculty of Health and Human Sciences, Plymouth University, Peninsula Allied Health Centre, Derriford Road, Plymouth, PL6 8BH, UK and Optometry Research Group, IOBA Eye Institute, School of Optometry (University of Valladolid), Valladolid

Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijo.IJO_266_18

Rights and Permissions
  Abstract 


Purpose: To determine repeatability and agreement of the ARK-30 handheld autorefractor with retinoscopy under cycloplegic and noncycloplegic conditions in children. Methods: Three consecutive autorefractor measurements (with and without cycloplegia) and retinoscopy were performed and compared in 30 randomized eyes of 30 children (mean age of 6.7 ± 2.7 years with spherical equivalent [SE] refraction from ‒4.01 to +7.38 D) in a cross-section and masked study. Bland–Altman analysis of autorefractor measurements (with and without cycloplegia) and agreement with retinoscopy were calculated with conventional notation (sphere [Sph] and cylinder [Cyl]) and vector notation (SE, J0, and J45coefficients). Results: ARK-30 measurements without cycloplegia were lower than under cycloplegic conditions (Sph: ‒0.52 ± 2.37 D vs + 0.86 ± 2.60 D, P < 0.01; Cyl: ‒0.83 ± 0.80 D versus ‒0.78 ± 0.77 D, P = 0.37; and SE: ‒0.94 ± 2.19 D vs + 0.47 ± 2.44 D, P < 0.01, respectively) and statistically different (P < 0.03) from retinoscopy (Shp: +0.83 ± 2.66 D; Cyl: ‒0.71 ± 0.87 D; SE: +0.51 ± 2.49 D). Without statistical differences were in J0and J45coefficients. Cyloplegic autorefraction measures were not found to be statistically significantly different to retinoscopy measures. ARK-30 under cycloplegia shows better repeatability with lower limits of agreement (LoA) in Sph (LoA: ‒0.66 to +0.69 D), and SE (LoA: ‒0.66 to +0.65 D) than without cycloplegia (LoA: ‒1.45 to +1.77 D, and ‒1.38 to +1.74 D, respectively). Conclusion: Under noncycloplegic conditions, ARK-30 autorefractor has low repeatability and a tendency toward minus over correction in children. However, repeatability and agreement with retinoscopy under cycloplegic conditions allow use of ARK-30 in children to estimate refraction but not to substitute gold standard retinoscopic refraction.

Keywords: Autorefraction, children refractive assessment, cycloplegia, repeatability, retinoscopy


How to cite this article:
Hernandez-Moreno L, Vallelado-Alvarez A, Martin R. Repeatability of ARK-30 in a pediatric population. Indian J Ophthalmol 2018;66:1262-7

How to cite this URL:
Hernandez-Moreno L, Vallelado-Alvarez A, Martin R. Repeatability of ARK-30 in a pediatric population. Indian J Ophthalmol [serial online] 2018 [cited 2024 Mar 29];66:1262-7. Available from: https://journals.lww.com/ijo/pages/default.aspx/text.asp?2018/66/9/1262/239343



Objective refraction, including retinoscopy and autorefraction, plays a crucial role to identify and correct refractive errors in children, helping to prevent and reduce the risk of amblyopia. Autorefraction has demonstrated the ability to give quick, repeatable, and accurate readings of refractive error in children without examiner bias.[1],[2] Nevertheless, for some instruments, pseudomyopia caused by accommodation and inadequate autofogging mechanisms have been reported.[3],[4],[5] Other factors such as fixation instabilities,[6] small pupils,[7] and media changes [8] can increase autorefractor variability in all subjects regardless of age. Therefore, retinoscopy, with and without cycloplegic is a necessary test in pediatric patients' eye exam.[9],[10]

Portable or handheld autorefractors have been proposed to improve pediatric patients' objective refraction and have been compared to table-mounted autorefractors, video-refraction, and retinoscopy.[2],[9],[11] ARK-30 handheld autorrefractor (Nidek Co. LTD, Aichi, Japan) is a portable autorefractor demonstrating good results after cataract surgery.[12] However, there are no previous reports that support its use in children population.

The aims of this study were (1) evaluate the repeatability of the ARK-30 handheld autorefractor under cycloplegic and noncycloplegic conditions and (2) determine the agreement between this instrument with retinoscopy in a pediatric sample.


  Methods Top


Study population

This was a cross-section and masked study. A baseline eye examination was performed on all subjects comprising visual acuity assessment, objective refraction techniques (with and without cycloplegia), binocular balancing, slit-lamp examination, and direct and indirect ophthalmoscopy. Subjects were excluded if they had significant pathology that could influence objective measurement of refraction such as congenital cataract or corneal leukomas affecting the visual axis. In order to achieve adequate cycloplegia autorefraction measures were conducted 30 min after the instillation the last drop of cyclopentolate hydrochloride 1% (Colircusi Cicloplejico, ALCON CUSI, Spain). Two drops in 10 min were administrated in the inferior cul de sac.

Written-informed consent was obtained from parents of each subject after approval of this study by the Human Sciences Ethics Committee of the University of Valladolid. All subjects were treated in accordance with the Declaration of Helsinki.

Refraction measures

The ARK-30 (Nidek Co. LTD, Aichi, Japan) is a portable autorefractor with a measurement range of ‒20.00 to +22.00 D for sphere and 12.00 D for cylinder. All subjects underwent autorefraction using the normal mode with the fogging mechanism activated to avoid instrument myopia or accommodation and to obtain a result comparable to a 6-m subjective refraction. In this mode, the “autoshot” facility permitted automated serial measurements when the instrument was in focus. Moreover, even ARK-30 autorefractor allows faster refractive power measurements (around 0.2 s); a minimum of child's collaboration that must look to device's target is required to achieve the autorefraction.

One trained optometrist performed three consecutive “autoshot” measurements in normal mode on each eye. The mean of these measurements was used as the final value for comparison to retinoscopy outcome. To avoid bias related with the use of both eyes of the same patient, data of one eye were randomized chosen to conduct the statistical comparisons.

To maintain masking, a different trained optometrist conducted cycloplegic retinoscopy (Heine Beta 200, Heine Optotechnik, Herrsching, Germany) and subjective refraction, when cooperation permitted.

Assessing the variance in the astigmatism poses a problem in the conventional clinical notation (e.g. ‒3.25 × 20 degrees). Therefore, the sphere, cylinder, and axis components were converted into a vector representation [13] with the coefficients M, J0, and J45 where M was the spherical equivalentlens of power equal to the mean spherical equivalence (M = sphere + [cylinder/2]), J0 was the Jackson cross-cylinder at axis 0° (J0= –[cylinder/2] cos[2 × axis]), and J45 was the Jackson cross-cylinder at axis 45° (J45= –[cylinder/2] sin[2 × axis]).

Statistical analysis

The data were analyzed using the Statistical Package for the Social Sciences (SPSS for Windows software, version 22.0, SPSS, Inc., Chicago, IL, USA). Normal distribution of variables was assessed using the Kolmogorov–Smirnov test (P values < 0.05 indicated that the data were not normally distributed). Results were presented as means ± standard deviation (SD), and 95% confidence interval (CI 95%).

Repeatability is the variation in measurements taken by the same operator with the same instrument on the same subject and under the same conditions, between multiple testings. The repeatability of ARK-30 was evaluated by obtaining three automated refraction measurements under cycloplegic and noncycloplegic conditions. The intraclass correlation coefficient (ICC) was calculated. The differences between each of the three test readings were averaged to determine the mean difference for sphere, cylinder, spherical equivalent, and the coefficients J0 and J45 with and without cycloplegia. Repeated analysis of variance (ANOVA) was used to detect differences between the three measures.

The difference between automated refraction with and without cycloplegia was assessed using Bland–Altman analysis.[14] The degree of agreement between automated refraction (with and without cycloplegia) and cycloplegic retinoscopy was, also, evaluated using Bland–Altman analysis.[14] The differences between two measurements were plotted against the averages. Limits of agreement were calculated (mean ± 1.96 SD). Significant differences were tested with Wilcoxon rank test. For all comparisons, P values of <0.05 were considered statistically significant. The relationship between mean value (x) and the difference (y) was determined using linear regression analyses, r2 correlation coefficient was calculated to test-retest reliability (P values of <0.05 were considered statistically significant).


  Results Top


The study comprised 30 randomized eyes (14 OD and 16 OS) of 30 children (18 boys, 12 girls) with a mean age of 6.7 ± 2.7 years (range: 3–13 years) with a mean spherical equivalent refraction (ARK-30 with cycloplegia) of ‒0.47 ± 2.44 D; (range: ‒4.01 D to +7.38 D). [Table 1] summarizes ARK-30 autorefraction (with and without cycloplegia) and cycloplegic retinoscopy results (sphere, cylinder, spherical equivalent, and J0 and J45 coefficients).
Table 1: Summary of the ARK-30 refraction without and under cycloplegia and retinoscopy

Click here to view


Repeatability of ARK-30 refraction

ARK-30 provided lower repeatable refraction measurements of sphere, cylinder, spherical equivalent, and J0 and J45 coefficients without cycloplegia than those achieved under cycloplegic conditions [Table 2].
Table 2: Summary of the ARK-30 repeatability refraction without and under cycloplegia

Click here to view


ARK-30 showed a tendency toward negative results when is used under noncycloplegic conditions [Figure 1] with a mean difference of ‒1.38 ± 1.19 D, LoA: ‒3.72 to +0.96 D, (P < 0.01); ICC = 0.950; r2 = 0.04, (P = 0.06) for sphere; a difference of ‒0.06 ± 0.36 D, LoA: ‒0.76 to + 0.64 D, (P = 0.37); ICC = 0.974; r2 = 0.01, (P = 0.47) for cylinder; a difference of ‒1.41 ± 1.18 D, LoA: ‒3.71 to +0.90 D, (P < 0.01); ICC = 0.943; r2 = 0.05, (P = 0.04) for spherical equivalence; a difference of ‒0.06 ± 0.46 D, LoA: ‒0.95 to +0.84 D, (P = 0.70); ICC = 0.515; r2 = 0.02, (P = 0.20) for J0 coefficient, and a difference of + 0.04 ± 0.56 D, LoA: ‒1.05 to +1.13 D, (P = 0.65); ICC = ‒0.147; r2 = 0.02, (P = 0.19) for J45 coefficient.
Figure 1: Bland–Altman plot comparing the difference between ARK-30 measurements collected under cycloplegic and under noncycloplegic conditions. Mean difference (continuous line) and limits of agreement (discontinuous line) were plotted to sphere (top-left), cylinder (top-right), spherical equivalence (bottom left), and J0 (bottom center) and J45 (bottom right) refraction coefficients

Click here to view


Agreement of ARK-30 automated refraction with retinoscopy

The differences between retinoscopy and ARK-30 (without cycloplegia) showed worse agreement and statistically significant differences in sphere (+1.37 ± 1.00 D, P < 0.01), cylinder (+0.15 ± 0.32 D, P = 0.03), and spherical equivalent (+1.45 ± 0.99 D, P < 0.01) [Table 3]. However, J0 and J45 coefficients showed nonstatistically significant differences. There were no apparent trends in the difference variabilities as a function of the mean values. The range of difference was from ‒0.50 (3.3%) to + 3.25 D (10%) in sphere and from ‒0.50 D (6.7%) to + 1.00 D (3.3%) in cylinder [Figure 2].
Table 3: Summary of the agreement between ARK-30 autorefraction and retinoscopy, without and under cycloplegia

Click here to view
Figure 2: Frequency of difference between the retinoscopy and ARK-30 under (a) and without (b) cycloplegia for sphere and cylinder. Positive differences mean than ARK-30 shows a higher myopic value

Click here to view


However, nonstatistically significant differences were found between retinoscopy and ARK-30 conducted under cycloplegia (P > 0.49) [Table 3]. The range of difference was from ‒2.75 (3.3%) to + 1.75 D (3.3%) in sphere and from ‒0.75 D (3.3%) to +0.75 D (3.3%) in cylinder [Figure 2]. There were no apparent trends in the difference variabilities as a function of the mean values.


  Discussion Top


Noncycloplegic autorefraction is a popular technique widely used to know the objective refractive status in children, conducted in several situations like vision screening, clinical practice, or in research settings, for example, in epidemiologic studies, clinical trials or others.[10]

The validity and repeatability of autorefraction have been widely studied in different populations: children,[9],[10],[11],[15],[16],[17],[18] youth,[6],[19] and adults,[4],[5],[20],[21],[22],[23],[24] using several devices.[25] However, to the best of our knowledge, this is the first report about the clinical application of the portable ARK-30 autorefractor in pediatric population (between 3 and 13 years old).

Our results are consistent with data from previous studies.[3],[4],[5],[11],[20],[21],[22],[23] We found high repeatability in sphere, cylinder, spherical equivalent, and J0 and J45 coefficients under cycloplegic and noncycloplegic conditions, with slightly better repeatability (lower limits of agreement [LoA]) when autorefraction was conducted under cycloplegia [Table 2].[9],[17]

We found good agreement between ARK-30 results under cycloplegic conditions and cycloplegic retinoscopy. Retinoscopy and subjective refraction are the gold standard in pediatric population assessments;[10],[11] however, good trained practitioner is required to achieve reliable retinoscopy results.[24] We found autorefractor values for sphere and spherical equivalent more negative (mean of 1.37 ± 1.00 D [ P < 0.01] and 1.45 ± 0.99 D [ P < 0.01], respectively [Table 3]) than cycloplegic retinoscopy (hyperopic underestimation or tendency toward myopic overcorrection), similar to previous reports.[9],[10],[11],[19] Differences are minimized when autorefraction was conducted under cycloplegia in sphere and spherical equivalent results (‒0.01 ± 0.86 D [P = 0.46] and + 0.04 ± 0.86 D [P = 0.38], respectively [Table 3]). Our results agree with previous reports that suggest than autorefractors without cycloplegia do not avoid accommodation; showing over-negative refractive outcome.[10],[11],[15],[17],[19] Harvey et al.[9] concluded that autorefraction (Nikon Retinomax) is reproducible and reliable in young children under cycloplegia as our results supports.[10]

Our results improve previous reports of other handheld autorefractometers in children. Suryakumar and Bobier [15] found a larger difference in comparison to retinoscopy refraction with a difference of + 0.37 ± 0.45 D under cycloplegic conditions and a difference of ‒1.15 ± 1.47 D without cycloplegia, in a child population (3–5 years) using three different devices (Retinomax K plus; Welch Allyn SureSight, and Power Refractor). Wesemann and Dick 17 reported a negative overcorrection higher of 2.0 D in almost one in four children between the age of 2 and 12 years when cycloplegia was not used, similar to the results of the present study. However, under cycloplegia we did not find any case with a negative overcorrection higher of 1.75 D [Figure 2].

Other studies, using different models of autorefractors, showed the same tendency of the autorefraction to underestimate refractive error relative to retinoscopy and subjective refraction.[4],[5],[21] Farook et al.[4] find 0.75 D over minus result with ‒2.20 to + 0.70 limits of agreement of Retinomax handheld autorefractor, compared with subjective refraction in an adults population (between 21 and 40 years). Zadnik et al.[24] and Rosenfield et al.[26] concluded than autorefractor (Canon R-1 autorefraction) is a valuable tool in adult refractive testing, having good correlation with values obtained after subjective refraction.

Autorefraction is not an accepted substitute for prescribing spectacles,[27] because refraction and prescribing are different concepts,[12] so autorefraction should not be perceived as substitute for the gold standard retinoscopic refraction. Binocular balancing, measurement of oculomotor coordination and accommodation assessment, are some aspects covered with subjective refraction that autorefraction cannot evaluate. However, our results suggest that the ARK-30 handheld autorefractor serves as a good tool to approximate children's refractive error. Furthermore, the differences between measures with and without cycloplegia,[28] suggest than noncycloplegic measurements must be interpreted with caution and cycloplegia is highly recommended to achieve reliable autorefraction in children. A subjective measure of refraction, where possible, dependent on age and cooperation, is also important to consider prior to prescribing a spectacle correction in children. The results of this study showed that the ARK-30 is as quick, accurate, and repeatable as other autorefractors currently available under noncycloplegic conditions,[3],[21] improving their results under cycloplegia, with the advantages to be a portable device. So, this instrument may be of great useful for refractive assessment in children, especially with cycloplegia.

Our study had some advantages compared to other studies, because we include a repeatability (using three different measurements) and cycloplegic retinoscopy comparison analysis, with a masked design. Moreover, only one experienced observer performed retinoscopy, and thus interobserver bias was controlled and minimized. Second, a trained observer conducted all autorefraction measurements without knowledge of the retinoscopy results (masked design). This methodology reduces bias in data collection and analysis. Finally, our study population involved only children, with a wide range of refractions (sphere from ‒4.25 to +7.75 D), providing better analysis of the use of ARK-30 in pediatric population. However, the major limitations of this study may be related with a relative small sample size, and the range of refractions included. So, more research with large sample to provide a separate analysis ranking children by age, level of cooperation; and assessing of high refractions could be necessaries, to show if repeatability and agreement with cycloplegic retinoscopy could be influenced by the child age, cooperation, or the amount of refraction.


  Conclusion Top


In conclusion, ARK-30 handheld autorefractor is repeatable to achieve objective refraction in children with more repeatable values under cycloplegic. Autorefraction should not be perceived as substitute for the gold standard retinoscopic refraction because, under noncycloplegic conditions, ARK-30 autorefractor has lower repeatability and a tendency toward minus over correction in children resulting in over diagnosis of myopia, suggesting that noncycloplegic measurements must be interpreted with caution and cycloplegia is highly recommended to achieve repeatable autorefraction results in children.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
El–Defrawy S, Clarke WN, Belec F, Pham B. Evaluation of a hand-held autorefractor in children younger than 6. J PediatrOphthalmol Strabismus 1998;35:107-9.  Back to cited text no. 1
    
2.
Tuncer I, Zengin MO, Karahan E. Comparison of the Retinomax hand-held autorefractor versus table-top autorefractor and retinoscopy. Int JOphthalmol 2014;7:491-5.  Back to cited text no. 2
    
3.
Davies LN, Mallen EA, Wolffsohn JS, Gilmartin B. Clinical evaluation of the Shin-Nippon NVision-K 5001/Grand Seiko WR-5100K autorefractor. Optom Vis Sci 2003;80:320-4.  Back to cited text no. 3
    
4.
Farook M, Venkatramani J, Gazzard G, Cheng A, Tan D, Saw S. Comparisons of the handheld autorefractor, table-mounted autorefractor, and subjective refraction in Singapore adults. Optom Vis Sci 2005;82:1066-70.  Back to cited text no. 4
    
5.
Gwiazda J, Weber C. Comparison of spherical equivalent refraction and astigmatism measured with three different models of autorefractors. Optom Vis Sci 2004;81:56-61.  Back to cited text no. 5
    
6.
Rugin A, Harris WF. Refractive variation during autorefraction: multivariate distribution of refractive status. Optom Vis Sci 1975;72:403-10.  Back to cited text no. 6
    
7.
Winn B, Pugh JR, Gilmartin B, Owens H. The effect of pupil size on static an dynamic measurements of accommodation using an infra–red optometer. Ophthalmic PhysiolOpt1989;9:277-83.  Back to cited text no. 7
    
8.
Elliott DB, Wilkes RD. A clinical evaluation of the Topcon RM6000 autorefractor. ClinExpOptom1989;72:150-3.  Back to cited text no. 8
    
9.
Harvey EM, Miller JM, Wagner LK, Dobson V. Reproducibility and accuracy of measurements with a hand held autorefractor in children. Br J Ophthalmol1997;81:941-8.  Back to cited text no. 9
    
10.
Choong YF, Chen AH, Goh PP. A comparison of autorefraction and subjective refraction with and without cycloplegia in primary school children. Am J Ophthalmol2006;142:68-74.  Back to cited text no. 10
    
11.
Oral Y, Gunaydin N, Ozgur O, Arsan A, Oskan S. A Comparison of Different Autorefractors With Retinoscopy in Children. J PediatrOphthalmol Strabismus 2012;49:370-7.  Back to cited text no. 11
    
12.
de Juan V, Herreras JM, Martin R, Morejon A, Perez I, Rio-San Cristobal A, Rodriguez G. Repeatability and agreement of ARK30 autorefraction after cataract surgery. ClinExpOphthalmol2012;40:134-40.  Back to cited text no. 12
    
13.
Thibos LN, Wheeler W, Horner D. Power vectors: An application of Fourier analysis to the description and statistical analysis of refractive error. Optom Vis Sci 1997;74:367-75.  Back to cited text no. 13
    
14.
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307-10.  Back to cited text no. 14
    
15.
Suryakumar R, Bobier WR. The manifestation of noncycloplegic refractive state in pre-school children is dependent on autorefractor design. Optometry Vis Sci 2007;80:578-86.  Back to cited text no. 15
    
16.
Virgili G, Angi M, Heede S, Rodriguez D, Bottega E, Molinari A. PoweRefractor versus Canon R–50 Autorefraction to assess refractive error in children: A community–based study in Ecuador. Optom Vis Sci 2007;84;144-8.  Back to cited text no. 16
    
17.
Wesemann W, Dick B. Accuracy and accommodation capability of a handheld autorefractor. J Cataract Refract Surg2000;26:62-70.  Back to cited text no. 17
    
18.
Cordonnier M, Dramaix M, Kallay O, de Bideran M. How accurate is the hand-held refractor Retinomax (R) in measuring cycloplegic refraction: A further evaluation. Strabismus 1998;6:133-42.  Back to cited text no. 18
    
19.
Kinge B, Midelfart A, Jacobsen G. Clinical evaluation of the Allergan Humphrey 500 autorefractor and the Nidek AR-1000 autorefractor. Br J Ophthalmol1996;80:35-9.  Back to cited text no. 19
    
20.
Allen PM, Radhakrishnan H, O'Leary DJ. Repeatability and validity of the PoweRefractor and the Nidek AR600-A in an adult population with healthy eyes. Optom Vis Sci 2003;80:245-51.  Back to cited text no. 20
    
21.
Jorge J, Queirós A, Almeida JB, Parafita MA. Retinoscopy/autorefraction: Which is the best starting point for a noncycloplegic refraction?. Optom Vis Sci 2005;82:64-8.  Back to cited text no. 21
    
22.
Cleary G, Spalton DJ, Patel PM, Lin PF, Marshall J. Diagnostic accuracy and variability of autorefraction by the Tracey Visual Function Analyzer and the Shin-Nippon NVision–K 5001 in relation to subjective refraction. OphtalmicPhysiolOpt2009;29:173-81.  Back to cited text no. 22
    
23.
Bullimore MA, Fusaro RE, Adams CW. The repeatability of automated and clinician refraction. Optom Vis Sci 1992;75:617-22.  Back to cited text no. 23
    
24.
Zadnik K, Mutti DO, Adams AJ. The repeatability of measurement of the ocular components. Invest Ophthalmol Vis Sci 1992;33:2325-33.  Back to cited text no. 24
    
25.
McCaghrey GE, Matthews FE. Clinical evaluation of a range of autorefractors. OphthalmolPhysiolOpt1993;13:129-37.  Back to cited text no. 25
    
26.
Rosenfield M, Chiu NN. Repeatability of subjective and objective refraction. Optom Vis Sci 1995;72:577-9.  Back to cited text no. 26
    
27.
Schmidt–Bacher AE, Kahlert C, Kolling G. Accuracy of two autorefractors–Pediatric Autorefractor Plusoptix and Retinomax–in cycloplejic children in comparison to retinoscopy. KlinMonblAugenheilkd2010;227:792-7.  Back to cited text no. 27
    
28.
Salvesen S, Kohler M. Automated refraction. A comparative study of automated refraction with the Nidek AR-1000 autorefractor and retinoscopy. Acta Ophthalmol1991;69:342-6.  Back to cited text no. 28
    


    Figures

  [Figure 1], [Figure 2]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]


This article has been cited by
1 Comparisons of objective and subjective refraction with and without cycloplegia using binocular wavefront optometer with autorefraction and retinoscopy in school-age children
Yadi Lei, Xun Chen, Mingrui Cheng, Boliang Li, Yinjie Jiang, Yilin Xu, Xiaoying Wang
Graefe's Archive for Clinical and Experimental Ophthalmology. 2022;
[Pubmed] | [DOI]
2 Comparative Study of Refraction between Wave Front-Based Refraction and Autorefraction without and with Cycloplegia in Children and Adolescents
Ana M Calvo-Maroto, Sara Llorente-González, Jaione Bezunartea-Bezunartea, Francisco Javier Hurtado-Ceña, Clara Berrozpe-Villabona, Valentina Bilbao-Malavé, David P Piñero, Jesús Barrio-Barrio, Sergio Recalde-Maestre
Children. 2022; 9(1): 88
[Pubmed] | [DOI]
3 Agreement and Repeatability of Central and Peripheral Refraction by One Novel Multispectral-Based Refractor
Weicong Lu, Rongyuan Ji, Wenzhi Ding, Yuyin Tian, Keli Long, Zhen Guo, Lin Leng
Frontiers in Medicine. 2021; 8
[Pubmed] | [DOI]
4

Comparison of Anterior Segment Parameters and Axial Length Using Two Scheimpflug Devices with Integrated Optical Biometers

Chantaka Supiyaphun, Sasivimol Rattanasiri, Passara Jongkhajornpong
Clinical Ophthalmology. 2020; Volume 14: 3487
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
Abstract
Methods
Results
Discussion
Conclusion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed1810    
    Printed31    
    Emailed0    
    PDF Downloaded209    
    Comments [Add]    
    Cited by others 4    

Recommend this journal