ORIGINAL ARTICLE
Year : 2020  |  Volume : 68  |  Issue : 6  |  Page : 1040-1043

Comparing the effect of three Descemet membrane endothelial keratoplasty injectors on endothelial damage of grafts


1 Gavin Herbert Eye Institute, and University of California, Irvine, Irvine, California
2 CorneaGen, Seattle, Washington, USA

Date of Submission17-Jul-2019
Date of Acceptance13-Jan-2020
Date of Web Publication25-May-2020

Correspondence Address:
Dr. Elizabeth Shen
Gavin Herbert Eye Institute, University of California Irvine, 850 Health Sciences Road, Irvine, CA 92697
California
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijo.IJO_1280_19

Rights and Permissions
  Abstract 


Purpose: Various injectors are commercially available for Descemet membrane endothelial keratoplasty (DMEK) but not all injectors have been studied for endothelial damage of grafts. The aim of the study was to compare endothelial damage in pre-stripped DMEK tissue from three clinically used injector devices: the modified Jones tube, the STAAR intraocular (IOL) injector, and the Geuder glass cannula in a laboratory setting. Methods: Twenty-four human donor corneas were used for this study, eight for each study arm. Each endothelial graft was pre-stripped, trephined to 8.0 mm diameter, then loaded into either the modified Jones tube, the STAAR IOL injector, or the Geuder glass cannula by an eye bank technician who had no prior experience with any of the injectors. Grafts were then ejected, stained with Calcein acetoxymethyl (AM), and quantitatively analyzed using FIJI image software. The primary outcome was the percent of endothelial damage from injector loading and injection. Donor demographics were analyzed using Fisher's exact test. The percentage of endothelial cell loss was compared across groups using the Kruskal–Wallis test. Results: The mean percent of endothelial damage from after injection of the graft was 37.8% (±SD 12.2%) for the modified Jones tube, 37.0% (±SD 13.9%) for the STAAR IOL injector, and 23.5% (±SD 5.1%) for the Geuder cannula (P = 0.008). Conclusion: DMEK injectors contribute to intraoperative endothelial damage of transplanted grafts. The Geuder glass cannula may offer increased ease of use and less endothelial damage compared to the modified Jones tube or STAAR IOL injector for the novice user in early cases.

Keywords: Descemet membrane endothelial keratoplasty, Descemet membrane endothelial keratoplasty injector, Geuder glass cannula, graft preparation, modified Jones tube


How to cite this article:
Shen E, Fox A, Johnson B, Farid M. Comparing the effect of three Descemet membrane endothelial keratoplasty injectors on endothelial damage of grafts. Indian J Ophthalmol 2020;68:1040-3

How to cite this URL:
Shen E, Fox A, Johnson B, Farid M. Comparing the effect of three Descemet membrane endothelial keratoplasty injectors on endothelial damage of grafts. Indian J Ophthalmol [serial online] 2020 [cited 2024 Mar 29];68:1040-3. Available from: https://journals.lww.com/ijo/pages/default.aspx/text.asp?2020/68/6/1040/284768



Descemet membrane endothelial keratoplasty (DMEK) has increased in popularity in recent years, largely due to rapid postoperative visual rehabilitation and excellent visual outcomes [1],[2],[3],[4],[5],[6] since its advent, numerous techniques and injection methods have been described.[7],[8],[9],[10],[11],[12] Commonly used injectors include the modified Jones tube (Gunther Weiss Scientific Glass, Portland, Oregon) and the STAAR intraocular lens (IOL) injectors, which are The Food and Drug Administration (FDA)-approved in the United States and used off-label for DMEK.

One of the primary goals in DMEK is the preservation of endothelial cells in the graft by minimizing trauma during manipulation of the issue. Endothelial cell loss may occur at all stages of the procedure, from eye bank preparation of the graft and loading of the tissue into the injector to insertion into the anterior chamber and even unfolding and centering the graft. The use of a closed-system injection system, such as the modified Jones tube or an IOL injector, allows the graft to be suspended in fluid and injected into the anterior chamber without direct contact with instrumentation. Prior studies comparing and quantifying endothelial cell loss between different injector methods have evaluated the modified Jones tube and IOL cartridges.[13],[14] The Geuder glass injector (Geuder AC, Heidelberg, Germany) was recently introduced for the purpose of DMEK insertion but there have been no studies to date on its effect on endothelial cells. In this study, we compare the effect of three different injectors—the modified Jones tube, the STAAR IOL injector (STAAR Surgical, Monrovia, California), and the Geuder glass cannula—on endothelial cell loss in DMEK grafts.


  Methods Top


Tissue preparation

Twenty-four corneas were prepared using the eye bank's DMEK protocol by one eye bank technician. Grafts were trephined to 8.0 mm, stained with Trypan blue (Invitrogen, Eugene, Oregon) for 30 s to visualize the edge of the tissue, and then suspended in a balanced salt solution (BSS) (Alcon Laboratories Inc., Fort Worth, Texas) until ready for aspiration into the injector. A prespecified number of eight grafts were used in testing for each injector [Figure 1].
Figure 1: Descemet membrane endothelial keratoplasty injector systems. (a) Modified Jones tube, (b) STAAR intraocular lens injector, (c) Geuder glass cannula

Click here to view


Of note, the eye bank technician did not have any prior experience with any one injector and received basic training on loading the grafts into each injector by the surgeon (M.F.) prior to the study. For the Jones tube, grafts were aspirated through the injection tip of the tube. For the STAAR IOL injector, grafts were loaded into the IOL cartridge under BSS, which was then placed into the injector. For the Geuder cannula, grafts were aspirated through the larger entry port of the cannula. For each injector, grafts were ejected into a 50 μg Calcein AM (Invitrogen, Eugene, Oregon)/100 μL dimethyl sulfoxide (Sigma-Aldrich, St Louis, Missouri)/20 mL BSS solution and were allowed to stain for 15 min. Grafts were then transferred to a flattened bed of Healon 5 (Johnson & Johnson Vision, Santa Ana, California) in a  Petri dish More Details and carefully unscrolled using a few drops of BSS, toothless forceps, and a Sinskey hook. Once unscrolled, the grafts were covered with a layer of Healon 5 for endothelium protection and for holding the graft in place.

Endothelial cell viability analysis

The stained grafts were then imaged using an inverted light microscope using blue light (495–515 nm) filter. Images were digitally stitched together using Microsoft Paint into a composite image. Viable cells appear as hyperfluorescent or bright green, whereas areas of cell death appear hypofluorescent or dark [Figure 2].
Figure 2: After injection of a graft with either (a) modified Jones tube, (c) STAAR intraocular lens injector, or (e) Geuder glass cannula, the tissue was stained with Calcein-acetoxymethyl, imaged using an inverted light microscope and stitched to create a composite image. Images were then segmented using FIJI software (b, d and f)

Click here to view


FIJI open software,[15] previously described as a reliable method for performing cell counts, was used to estimate endothelial cell viability.[16] The individual performing the analysis of endothelial cell viability was not the technician who performed the experiment and was blinded to the injector used.

Statistical methods

Our study was powered to detect a 10% difference in endothelial cell loss between three injector groups with a confidence level of 90% and an α of 0.05. This resulted in eight grafts per group. Donor demographics were analyzed using Fisher's exact test. The amount of endothelial cell loss was estimated as a percentage and compared across groups using the Kruskal–Wallis test. Analyses were performed using STATA (Version 13.0).


  Results Top


Twenty-four grafts were prepared for this study, eight in each arm. There were no significant differences in donor age or death-to-preservation time between groups, though endothelial cell counts were lower in the Geuder glass cannula group (P = 0.02) [Table 1].
Table 1: DMEK* graft demographics

Click here to view


Grafts injected with the modified Jones tube had a mean percentage of endothelial cell loss of 37.8% (±SD 12.2%). Grafts injected with the STAAR IOL injector had a mean percentage cell loss of 37.0% (±SD 13.9%). Grafts injected with the Geuder glass cannula had a mean percentage cell loss of 23.5% (±SD 5.1%), which was a statistically significant difference compared to the other two injector methods (P = 0.008, Kruskal–Wallis test) [Table 2] and [Figure 3]. Both the modified Jones tube group and STAAR IOL injector group had one outlier each: 64.9% and 65.3% endothelial cell loss, respectively, due to stiff tissue that did not scroll. The mean percentage endothelial cell loss when these two outliers were excluded was 34.9% (±SD 7.4%) in the modified Jones tube group and 32.9% (±SD 8.6%) in the STAAR IOL injector group. The difference in mean percentage endothelial cell loss across the three groups remained statistically significant when the outliers were excluded from the analysis (P = 0.01).
Table 2: Comparison of endothelial cell loss after graft preparation and injection

Click here to view
Figure 3: Box-plot comparison of endothelial cell loss after graft preparation and injection

Click here to view



  Discussion Top


In this study, we compared the amount of endothelial cell damage induced by three clinically used DMEK injectors. We found the Geuder glass cannula caused less damage than either the modified Jones tube or the STAAR IOL injector in the early learning curve of adopting these injectors.

Since the advent of DMEK, there has been continued refinement in technique. Different studies have proposed various injector techniques and expanded injector options.[2],[7],[8],[9],[10],[11],[12] One of the earliest injectors employed was the glass Pasteur pipette. In the USA today, commonly used injectors include the modified Jones tube, IOL injectors, and most recently, the Geuder glass cannula. The DORC injector (DORC, Zuidland, the Netherlands) is another option that is currently available only in Europe. The DORC and Geuder injectors are double port injectors specifically designed for DMEK use, such that there is a wide entry port for aspiration and a narrow exit port (sized for the corneal incision) for injection.

The primary differences in injectors reside in their material (i.e., glass or plastic) and the diameters of their openings. Prior studies have postulated that glass may be less toxic to endothelium compared to plastic, which may accidentally adhere to the endothelium.[8],[10],[17]

The diameter of the injector opening would intuitively appear to play a role, as smaller openings are more likely to traumatize the scrolled tissue during aspiration. However, existing studies have not supported this hypothesis. One study compared glass injectors with various openings (0.5 mm, 0.9 mm, and 1.4 mm) and did not find a statistically significant difference in cell viability.[17] Using the same technique as described in this study, Schallhorn et al. compared the modified Jones tube, a glass injector with a 2.4 mm opening, to the Viscoject IOL injector (Viscoject 2.2, Medicel), a plastic injector with a 1.9 mm opening, and found no statistically significant difference in cell viability.[14] More recently, Downes et al. compared the DORC injector with a 1.4 mm opening to the modified Jones tube and found no difference in cell loss.[18]

This is the first study to directly compare the Geuder glass cannula with two commercially available injectors in clinical use. We found a difference of 14.3% and 13.5% less endothelial damage caused by the Geuder glass cannula compared to the modified Jones tube and STAAR IOL injector, respectively, in a novice user.

Cell loss from graft preparation alone has been estimated to range from 12.4% to 28%, while tissue preparation and subsequent aspiration and injection can result in cell loss of 23% to 32%.[14],[18],[19],[20] Our results for the Geuder glass cannula are at least comparable; for the modified Jones tube, our reported cell loss is slightly higher than the 19–32% of endothelial loss reported in the literature.[14],[18],[21],[22],[23]

One possible explanation for our results compared to prior studies is that familiarity with the DMEK injection technique has been shown to influence the amount of endothelial cell loss.[24] The abovementioned studies were performed by surgeons experienced with the injection technique; in our study, an eye bank technician who was naïve to the three injectors performed the stripping and injection of the grafts. The lack of prior experience likely resulted in a higher rate than expected of endothelial loss with the modified Jones tube compared with experienced users. However, this provides insight into the learning curve for each injector. It is plausible that the modified Jones tube and the STAAR IOL injector have a more difficult learning curve than the Geuder cannula, resulting in a higher endothelial loss in early cases. This suggests that the Geuder glass cannula may be easier to learn and be more forgiving of endothelial trauma.

One important limitation of this study is that our findings are limited to the laboratory setting and may not correlate with clinical outcomes. Droutsas et al. compared three glass injectors (DORC injector, Geuder glass cannula, and Pasteur pipette) in patients who underwent DMEK and did not find a clinically significant difference in endothelial cell density at 12-month postoperatively.[13] Future studies are needed to determine whether differences in these commonly used injector systems translate to a difference in the longevity of DMEK grafts implanted in patients. Another limitation to consider is that, while the technician lacked prior experience with the injector at the beginning of the experiment, the experience was gained throughout the study. The increasing familiarity with the injectors may have an undetermined effect on our results. We hypothesize that increasing experience would most affect the outcome of endothelial loss on injectors with the easiest learning curves; however, our sample size was not sufficient to assess for this effect.


  Conclusion Top


In conclusion, we found that the modified Jones tube, STAAR IOL injector, and Geuder glass cannula all offer acceptable tissue quality and cell loss for DMEK transplantation. We found statistically significant less cell loss induced by the Geuder glass cannula compared to the other two injectors, suggesting the Geuder glass cannula may have an easier learning curve for the novice user in early cases.

Acknowledgements

The authors thank CorneaGen for providing donor corneal tissue for the study.

Financial support and sponsorship

The authors have no funding sources to disclose. Tissue was donated by CorneaGen.

Conflicts of interest

Elizabeth Shen, MD has no proprietary interests and no financial disclosures. Adam Fox, CEBT is a former employee of CorneaGen. Brian Johnson, CEBT is an employee of CorneaGen. Marjan Farid, MD is a consultant for CorneaGen. No authors have any proprietary interests.



 
  References Top

1.
Price MO, Price FW. Descemet's membrane endothelial keratoplasty surgery: Update on the evidence and hurdles to acceptance. Curr Opin Ophthalmol 2013;24:329-35.  Back to cited text no. 1
    
2.
Price MO, Giebel AW, Fairchild KM, Price FW Jr. Descemet's membrane endothelial keratoplasty: Prospective multicenter study of visual and refractive outcomes and endothelial survival. Ophthalmology 2009;116:2361-8.  Back to cited text no. 2
    
3.
Ham L, Balachandran C, Verschoor CA, van der Wees J, Melles GR. Visual rehabilitation rate after isolated Descemet membrane transplantation: Descemet membrane endothelial keratoplasty. Arch Ophthalmol 2009;127:252-5.  Back to cited text no. 3
    
4.
Melles GR, Ong TS, Ververs B, van der Wees J. Descemet membrane endothelial keratoplasty (DMEK). Cornea 2006;25:987-90.  Back to cited text no. 4
    
5.
Deng SX, Sanchez PJ, Chen L. Clinical outcomes of Descemet membrane endothelial keratoplasty using eye bank-prepared tissues. Am J Ophthalmol 2015;159:590-6.  Back to cited text no. 5
    
6.
Oellerich S, Baydoun L, Peraza-Nieves J, Ilyas A, Frank L, Binder PS, et al. Multicenter study of 6-month clinical outcomes after Descemet membrane endothelial keratoplasty. Cornea 2017;36:1467-76.  Back to cited text no. 6
    
7.
Kruse FE, Laaser K, Cursiefen C, Heindl LM, Schlötzer-Schrehardt U, Riss S, et al. A stepwise approach to donor preparation and insertion increases safety and outcome of Descemet membrane endothelial keratoplasty. Cornea 2011;30:580-7.  Back to cited text no. 7
    
8.
Dapena I, Moutsouris K, Droutsas K, Ham L, van Dijk K, Melles GR. Standardized “no-touch” technique for Descemet membrane endothelial keratoplasty. Arch Ophthalmol 2011;129:88-94.  Back to cited text no. 8
    
9.
Kim EC, Bonfadini G, Todd L, Zhu A, Jun AS. Simple, inexpensive, and effective injector for Descemet membrane endothelial keratoplasty. Cornea 2014;33:649-52.  Back to cited text no. 9
    
10.
Terry MA, Straiko MD, Veldman PB, Talajic JC, VanZyl C, Sales CS, et al. Standardized DMEK technique: Reducing complications using prestripped tissue, novel glass injector, and sulfur hexafluoride (SF6) gas. Cornea 2015;34:845-52.  Back to cited text no. 10
    
11.
Bahar I, Kaiserman I, Sansanayudh W, Levinger E, Rootman DS. Busin guide vs forceps for the insertion of the donor lenticule in Descemet stripping automated endothelial keratoplasty. Am J Ophthalmol 2009;147:220-6.e1.  Back to cited text no. 11
    
12.
Arnalich-Montiel F, Munoz-Negrete FJ, De Miguel MP. Double port injector device to reduce endothelial damage in DMEK. Eye (Lond) 2014;28:748-51.  Back to cited text no. 12
    
13.
Droutsas K, Lazaridis A, Kymionis GD, Chatzistefanou K, Moschos MM, Koutsandrea C, et al. Comparison of endothelial cell loss and complications following DMEK with the use of three different graft injectors. Eye (Lond) 2018;32:19-25.  Back to cited text no. 13
    
14.
Schallhorn JM, Holiman JD, Stoeger CG, Chamberlain W. Quantification and patterns of endothelial cell loss due to eye bank preparation and injector method in Descemet membrane endothelial keratoplasty tissues. Cornea 2016;35:377-82.  Back to cited text no. 14
    
15.
Schindelin J, Arganda-Carreras I, Frise E, Kaynig V, Longair M, Pietzsch T, et al. Fiji: An open-source platform for biological-image analysis. Nat Methods 2012;9:676-82.  Back to cited text no. 15
    
16.
Jardine GJ, Holiman JD, Stoeger CG, Chamberlain WD. Imaging and quantification of endothelial cell loss in eye bank prepared DMEK grafts using trainable segmentation software. Curr Eye Res 2014;39:894-901.  Back to cited text no. 16
    
17.
Yoeruek E, Bartz-Schmidt KU, Hofmann J. Impact of the radius of the injector system on the cell viability in Descemet membrane endothelial keratoplasty: An ex vivo experimental study. Acta Ophthalmol 2016;94:e1-5.  Back to cited text no. 17
    
18.
Downes K, Tran KD, Stoeger CG, Chamberlain W. Cumulative endothelial cell loss in Descemet membrane endothelial keratoplasty grafts from preparation through insertion with glass injectors. Cornea 2018;37:698-704.  Back to cited text no. 18
    
19.
Altaan SL, Gupta A, Sidney LE, Elalfy MS, Agarwal A, Dua HS. Endothelial cell loss following tissue harvesting by pneumodissection for endothelial keratoplasty: An ex vivo study. Br J Ophthalmol 2015;99:710-3.  Back to cited text no. 19
    
20.
Salvalaio G, Parekh M, Ruzza A, Ferrari S, Camposampiero D, Ponzin D. DMEK lenticule preparation from donor corneas using a novel 'SubHyS' technique followed by anterior corneal dissection. Br J Ophthalmol 2014;98:1120-5.  Back to cited text no. 20
    
21.
Zeidenweber DA, Tran KD, Sales CS, Wehrer SW, Straiko MD, Terry MA. Prestained and preloaded DMEK grafts: An evaluation of tissue quality and stain retention. Cornea 2017;36:1402-7.  Back to cited text no. 21
    
22.
Wolle MA, DeMill DL, Johnson L, Lentz SI, Woodward MA, Mian SI. Quantitative analysis of endothelial cell loss in preloaded Descemet membrane endothelial keratoplasty grafts. Cornea 2017;36:1295-301.  Back to cited text no. 22
    
23.
Hamzaoglu EC, Straiko MD, Mayko ZM, Sáles CS, Terry MA. The first 100 eyes of standardized Descemet stripping automated endothelial keratoplasty versus standardized Descemet membrane endothelial keratoplasty. Ophthalmology 2015;122:2193-9.  Back to cited text no. 23
    
24.
Tran KD, Alabi RO, Odell K, Dye PK, Downes K, Sales CS. Measuring endothelial cell loss on DMEK grafts after transplantation in human cadaveric whole eyes: Description of the technique and pilot study. Cornea 2018;37:1075-80.  Back to cited text no. 24
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1], [Table 2]


This article has been cited by
1 Validation of preloaded DMEK donor tissues: a laboratory-based study on endothelial cell viability and comparison of two F-mark inks
Narisa Dhupar, Sara AlShaker, Jenny Chai, Judy Yan, Ness Little, Christine Humphreys, Clara C. Chan, Stephan Ong Tone
Canadian Journal of Ophthalmology. 2024;
[Pubmed] | [DOI]
2 Effect of Low-Temperature Preservation in Optisol-GS on Preloaded, Endothelium-Out DMEK Grafts
Alessandro Ruzza, Stefano Ferrari, Matteo Airaldi, Vito Romano, Diego Ponzin
Journal of Clinical Medicine. 2023; 12(3): 1026
[Pubmed] | [DOI]
3 Delivering Endothelial Keratoplasty Grafts: Modern Day Transplant Devices
Mohit Parekh, Vito Romano, Kareem Hassanin, Valeria Testa, Rintra Wongvisavavit, Stefano Ferrari, Colin Willoughby, Diego Ponzin, Vishal Jhanji, Namrata Sharma, Julie Daniels, Stephen B. Kaye, Sajjad Ahmad, Hannah Levis
Current Eye Research. 2022; : 1
[Pubmed] | [DOI]
4 Review: Optimizing DMEK Technique, 2018-2020
David A. Ramirez, Gregory A. Schmidt, Kristan S. Worthington, Christopher S. Sales
International Ophthalmology Clinics. 2021; 61(1): 45
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
    Similar in PUBMED
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
Abstract
Methods
Results
Discussion
Conclusion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed1334    
    Printed72    
    Emailed0    
    PDF Downloaded151    
    Comments [Add]    
    Cited by others 4    

Recommend this journal