Indian Journal of Ophthalmology

ORIGINAL ARTICLE
Year
: 2015  |  Volume : 63  |  Issue : 4  |  Page : 323--326

Comparison of intraocular pressure measurement with Scheimpflug-based noncontact tonometer with and without hydrogel contact lenses


Mukesh Kumar1, Rohit Shetty1, Chaitra Jayadev1, Debarun Dutta2, Maneck D Nicolsan1, Sriharsha Nagaraj1, Rajesh S Kumar1 
1 Narayana Nethralaya, Bengaluru, Karnataka, India
2 School of Optometry and Vision Science, The University of New Wales, Sydney, Australia

Correspondence Address:
Mukesh Kumar
Narayana Nethralaya, 121/C, Chord Road, 1st �DQ�R�DQ� Block, Rajajinagar, Bengaluru - 560 010, Karnataka
India

Objectives: The objective was to determine the repeatability of intraocular pressure (IOP) measurements made through a soft contact lens (CL) using the Scheimpflug noncontact tonometry in healthy subjects. Methods: This prospective, randomized, single-center study included one eye of 88 subjects (40 male and 48 female). Only participants without glaucoma or any other ocular pathology were included in this study. Three consecutive IOP measurements by the Scheimpflug noncontact tonometry were performed with and without daily disposable hydrogel CLs (−0.50 DS) (Dailies-nelfilcon A, 69% water, 8.7 mm base curve, 14 mm diameter, center thickness 0.10 mm) by a single operator. To avoid any bias arising from diurnal variation, all measurements were made at a similar time of day (11 am ± 1 h). The repeatability of IOP measurements using the Scheimpflug noncontact tonometry with and without CLs was evaluated using Pearson�SQ�s correlation analysis. Bland-Altman plotting was used to assess the limits of agreement between the measurements with and without CLs. Results: The mean (± standard deviation) IOPs with and without CL were 13.80 ± 2.70 and 13.79 ± 2.54 mm of Hg respectively. The mean difference was 0.01 ± 0.16 (95% confidence interval, +1.97 to − 2.00) mm Hg. Statistical analysis via paired t-test showed no statistical difference between the two groups with (P = 0.15). A good correlation was found for IOP measurements with and without CL (r = 0.93, P < 0.001). Good test-retest reliability was found when IOP was measured with and without CL. Conclusion: There was no significant difference between IOP measured with and without CLs by Scheimpflug noncontact tonometry.


How to cite this article:
Kumar M, Shetty R, Jayadev C, Dutta D, Nicolsan MD, Nagaraj S, Kumar RS. Comparison of intraocular pressure measurement with Scheimpflug-based noncontact tonometer with and without hydrogel contact lenses.Indian J Ophthalmol 2015;63:323-326


How to cite this URL:
Kumar M, Shetty R, Jayadev C, Dutta D, Nicolsan MD, Nagaraj S, Kumar RS. Comparison of intraocular pressure measurement with Scheimpflug-based noncontact tonometer with and without hydrogel contact lenses. Indian J Ophthalmol [serial online] 2015 [cited 2024 Mar 29 ];63:323-326
Available from: https://journals.lww.com/ijo/pages/default.aspx/article.asp?issn=0301-4738;year=2015;volume=63;issue=4;spage=323;epage=326;aulast=Kumar;type=0