Close | |
Figure 2: (a-f). Bland-Altman plot showing the agreement between measurements of central corneal thickness (CCT), anterior chamber depth (ACD), and nasal (N) angle opening distance (AOD) at 500 μm and 750 μm, with Casia and Visante ASOCT. The limits of agreement between Casia and Visante for ACD was +0.22 to −0.20 mm, CCT was +37.5 to −71.5 μm, N AOD 500 was +0.41 to −0.30, N AOD 750 was +0.50 to −0.34, N TISA 500 +0.19 to −0.12, and N TISA 750 was +0.27 to −0.19 |
|