Tonometry in normal and scarred corneas, and in postkeratoplasty eyes : A comparative study of the Goldmann, the ProTon and the Schiotz tonometers.


Department of Ophthalmology, Post-Graduate Institute of Medical Education and Research [PGIMER], Chandigarh, India

Correspondence Address:
A K Jain
Department of Ophthalmology, Post-Graduate Institute of Medical Education and Research [PGIMER], Chandigarh
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


PMID: 11271930

Rights and PermissionsRights and Permissions

PURPOSE: Clinical comparison of intraocular pressure (IOP) measured with the Goldmann applanation tonometer (GAT), the ProTon tonometer (PT), and the Schiotz tonometer (ST), in normal eyes, eyes with scarred corneas and postkeratoplasty eyes. MATERIAL AND METHODS: The IOP readings with GAT, PT, and ST were compared in 125 eyes with normal corneas (Group A), 17 eyes with scarred corneas (Group B), and in 21 postkeratoplasty eyes (Group C). The data were statistically analysed at 95% confidence interval; linear regression analysis and paired t-test were done. RESULTS: The mean differences and their standard deviation [SD] between GAT and PT readings, and GAT and ST readings respectively were: [1] in Group A: -0.23 [SD 2.75] mmHg and +0.24 [SD 3.18] mmHg respectively; [2] in Group B: -1.8 [SD 12.67] mmHg and -4.5 [SD 9.95 mmHg; and [3] in Group C: +0.24 [SD 8.72] mmHg and -0.12 [SD 8.7] mmHg. They were not statistically significant. In Group A the 95% confidence interval between GAT and PT readings was -5.27 mmHg to 5.73 mmHg, and between GAT and ST readings, -6.12 mmHg to 6.59 mmHg. Ninety six [77%] eyes with the PT and 84 [69%] eyes with ST measurements were within 3 mmHg of GAT pressure. The correlation coefficients [r] for PT and ST were 0.93 [P = 0.0000] and 0.88 [P = 0.0000] respectively. In Group B 95% confidence interval between GAT and PT readings was -27.17 mmHg to 23.51 mmHg, and between GAT and ST measurement, -24.37 mmHg to 15.44 mmHg. The correlation coefficients [r] for the PT and ST were 0.112 [P = 0.660] and 0.630 [P = 0.006] respectively. In group C, the 95% confidence interval between GAT and PT measurements was -17.20 mmHg to 17.67 mmHg, and between GAT and ST measurements, -17.51 mmHg to 17.27 mmHg. The correlation coefficients [r] for the PT and the ST were 0.780 [P = 0.0000] and 0.740 [P = 0.0001] respectively. CONCLUSIONS: In clinical practice PT appears to have a higher level of accuracy than ST in normal corneas. In scarred corneas and post-penetrating keratoplasty eyes, because of high SD for mean differences and wide confidence interval of 95%, both PT and ST are inaccurate in measuring IOP as compared to GAT in such eyes.


[FULL TEXT] [PDF Not available]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed8602    
    Printed234    
    Emailed6    
    PDF Downloaded1    
    Comments [Add]    
    Cited by others 5    

Recommend this journal