ORIGINAL ARTICLE
Year : 2006  |  Volume : 54  |  Issue : 4  |  Page : 261-265

Inter-instrument agreement and influence of central corneal thickness on measurements with Goldmann, pneumotonometer and noncontact tonometer in glaucomatous eyes


Dr. Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences, New Delhi - 110 029, India

Correspondence Address:
Viney Gupta
Dr. Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences, New Delhi - 110 029
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0301-4738.27952

Rights and Permissions

Purpose: This study was conducted to compare the intraocular pressure (IOP) measurements by the Goldman applanation tonometer (GAT), non contact tonometer (NCT) and the ocular blood flow (OBF) pneumotonometer in different IOP ranges in glaucomatous eyes. The effect of central corneal thickness (CCT) on IOP measurement in chronic glaucomatous eyes using the three different tonometers was also evaluated. Materials and Methods: IOP measurements of 130 eyes of primary glaucoma patients were performed using GAT by an ophthalmologist while NCT and OBF-pneumotonometer measurements were performed by an experienced optometrist. The IOP values were compared amongst the three instruments in the three different IOP ranges (0-18 mmHg, >18 to 25 mmHg, > 25 mmHg). CCT was also measured in all patients. Results: The mean of paired difference between GAT and NCT was 0.9 ± 3.1 mmHg while that between GAT and OBF-pneumotonometer was 0.3 ± 3.4 mmHg. The OBF-pneumotonometer and NCT were more affected by corneal thickness (0.41 mmHg and 0.4 mmHg / 10 µ corneal thickness respectively) while GAT was the least affected by corneal thickness (0.3 mmHg / 10 µ corneal thickness) though the difference was not statistically significant ( P =0.42). Conclusion: With appropriate correction for corneal thickness the NCT and OBF-pneumotonometer can be used as reliably as GAT in following up glaucomatous patients.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed6753    
    Printed160    
    Emailed9    
    PDF Downloaded581    
    Comments [Add]    
    Cited by others 21    

Recommend this journal