ORIGINAL ARTICLE
Year : 2009  |  Volume : 57  |  Issue : 5  |  Page : 361-364

Clinical measurement and categorization of optic disc in glaucoma patients


VST Glaucoma Centre, Kallam Anji Reddy Campus, L V Prasad Eye Institute, L V Prasad Marg, Banjara Hills, Hyderabad, India

Date of Submission11-Jul-2008
Date of Acceptance22-Dec-2008
Date of Web Publication14-Aug-2009

Correspondence Address:
Harsha B L Rao
VST Glaucoma Centre, Kallam Anji Reddy, Campus, L V Prasad Eye Institute, L V Prasad Marg, Banjara Hills, Hyderabad - 500 034
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0301-4738.55075

Rights and Permissions
  Abstract 

Background: Assessment of optic disc size is an important component of optic nerve head examination. Agreement between different methods of disc size measurements is not very good.
Purpose: To assess the agreement between the disc size assessed by Heidelberg retina tomograph (HRT) and stereobiomicroscopy with a 90 diopter (D) lens. To report the clinical (measured by biomicroscopy) disc diameters of small, average and large optic discs categorized by HRT disc areas.
Setting and Design: Observational study of subjects examined in the glaucoma clinic of a tertiary eye institute.
Materials and Methods: Seventy-five eyes of 75 glaucoma subjects were studied. Disc diameter was measured using stereobiomicroscopy and HRT. The agreement between the two sets of measurements was assessed by intraclass correlation coefficient (ICC). Discs were classified into small (<1.6 mm 2 ), average (1.6-2.6 mm 2 ) and large (>2.6 mm 2 ) depending on cutoffs provided by the manufacturers of HRT. The means (95% CI) of the corresponding vertical disc diameter in these groups were assessed.
Statistical Analysis: ICC, Bland and Altman plots.
Results: ICC for measurements of clinical and HRT horizontal disc diameter was 0.518 and for vertical disc diameter measurement was 0.487. The mean difference between the clinical and HRT measurements as analyzed by the Bland and Altman plot was 0.17 (95% CI, 0.13- 0.47) for horizontal and 0.22 (95% CI, 0.11- 0.54) for vertical disc diameter. Of the 75 eyes, 3 eyes had small discs, 54 average and 18 large discs. The mean clinical vertical disc diameter for small discs was 1.55 mm (95% CI, 1.2-1.7), for average discs was 1.91 mm (95% CI, 1.87-1.96) and for large discs was 2.15 mm (95% CI, 2.03-2.27).
Conclusion: The agreement between clinical and HRT disc diameter measurements is moderate. Disc diameter measurement on stereobiomicroscopy can be used to categorize discs into small, average and large discs.

Keywords: Heidelberg retinal tomograph


How to cite this article:
Rao HB, Sekhar G C, Babu GJ, Parikh RS. Clinical measurement and categorization of optic disc in glaucoma patients. Indian J Ophthalmol 2009;57:361-4

How to cite this URL:
Rao HB, Sekhar G C, Babu GJ, Parikh RS. Clinical measurement and categorization of optic disc in glaucoma patients. Indian J Ophthalmol [serial online] 2009 [cited 2024 Mar 29];57:361-4. Available from: https://journals.lww.com/ijo/pages/default.aspx/text.asp?2009/57/5/361/55075

Assessment of optic disc size is an important component of optic nerve head examination, as optic disc parameters such as neuroretinal rim area and cup-disc ratio vary with the disc size. The Blue Mountain study has shown that cup-disc ratio is strongly associated with disc diameter and optic discs with larger vertical diameters have considerably greater vertical cup-disc ratios. [1] Correction for disc size is necessary for optic disc variables when ranking these variables for detection of glaucomatous optic nerve damage. [2] It is debatable whether disc size is an independent risk factor for glaucoma. [3] Disc size is known to vary largely between populations, among individuals and between eyes. [4],[5] Differences in the techniques of measuring disc size across studies could result in different estimates of disc size, making comparisons among studies [6],[7] difficult. These differences must be considered while evaluating the optic disc. In practice it is not necessary to actually measure the disc size but is enough to classify it as large, medium or small disc on clinical examination. [8] We evaluated the disc diameter in the clinic by slit lamp biomicroscopy with a +90 diopter (D) lens and looked at its agreement with measurements taken using confocal scanning laser ophthalmoscopy. The aim of the study was to correlate the disc diameter assessed by biomicroscopy of small, average and large discs as categorization by Heidelberg retina tomography (HRT).


  Materials and Methods Top


This is an observational study of 75 eyes of 75 non-consecutive subjects attending the glaucoma clinic at our institute between January 2007 and March 2007. The ethics committee of the L.V. Prasad Eye Institute, Hyderabad approved the study protocol. The methods applied in the study adhered to the tenets of the declaration of Helsinki for the use of human subjects in biomedical research.

Inclusion criteria were age above 18 years, best corrected visual acuity of 20/40 or better, good images on HRT as defined by interscan standard deviation (SD) of ≤50 µm and willingness to participate in the study. The exclusion criteria were refractive errors exceeding 5.0 D sphere and/or 3.0 D cylinder, any media opacity precluding imaging techniques or clinical examination of the disc and inability to undergo the tests.

All subjects underwent a comprehensive ophthalmic examination. The vertical and horizontal disc diameters were recorded after mydriasis, at the slit lamp with a +90 D double-aspheric fundus lens (Volk Opticals, Mentor, OH), by a single observer (HBL). A narrow slit beam of the Haag-Streit slit lamp (Haag-Streit BM 900® V, Haag- Streit AG, Koeniz, Switzerland), with its width maintained constant, was progressively reduced in size from 8 mm until it coincided with the diameter of the disc. The beam length was then recorded from the millimeter scale of the instrument. Because the slit lamp beam length is calibrated to 0.1 mm, the reading was approximated to the nearest 0.1 mm. After each reading, the millimeter scale was reset to 8 mm. The length of the beam was adjusted first to the vertical and then to the horizontal diameter of the disc and read on the scale of the slit-lamp in millimeters. The measurement was taken only if the optic disc was in good focus and was seen in the center of the image field. This reading was then multiplied with a correction factor of 1.41 to obtain the actual disc diameter. Direct measurements from the slit lamp of small, average and large disc groups are also reported.

The intra- and inter-observer agreements for disc diameter measurements were studied in the initial 40 subjects who fulfilled the inclusion and the exclusion criteria. These subjects were not included in the study cohort of 75 subjects. Intra-observer agreement was assessed as follows. The observer (HBL) aligned the height of the slit beam to the diameter of the disc and the reading on the slit lamp scale was noted by an assistant. The observer was masked to the reading. After measuring the vertical and horizontal disc diameters of both the eyes, the slit beam was opened to full height and the disc diameters of the first eye was once again measured by aligning the height of the slit beam to the diameter of the disc and the reading on the slit lamp scale was noted by a second assistant who was masked to the reading noted down by the first assistant. The observer (HBL) was masked to both these measurements. Inter-observer agreement was assessed between the first set of readings of the first observer (HBL) and the disc diameter measurements obtained by an experienced observer (GCS) on the same 40 subjects, to validate the measurements of the first observer.

HRT-II (Heidelberg Engineering, GmbH, Dossenheim, Germany) examination was performed after entering the subject's manifest refraction. Three scans were acquired automatically after initial positioning by the operator. All scans were assessed subjectively for the presence of good optic nerve head (ONH) centration, focus and uniform illumination. All scans had interscan standard deviation of < 50 µ m. Scans with poor image quality were excluded. An experienced operator (GJB), masked to the clinical measurements, marked the optic disc margin as the inner border of the Elschnig ring. Contour lines were placed on the margin of the optic disc according to the instructions provided on the HRT Web site ( http://www.heidelbergengineering.com ). HRT disc area was recorded. Vertical and horizontal disc diameters were calculated using the interactive measurements option.

One randomly selected eye of each subject was chosen for analysis. The agreement between clinical and HRT disc diameter measurements was analyzed by intraclass correlation coefficient (ICC) and the Bland and Altman method. SPSS version 13 and Medcalc version 9 were used for statistical analyses.

Based on the standard deviation of a mean difference of 0.5, to estimate an upper agreement limit with an absolute error of 0.2 at 95% confidence level, 73 subjects were required.

The discs were then classified into small, average and large depending on the disc area cut-offs provided by the manufacturers of HRT. [9] Discs with area less than 1.6 mm 2 were classified as small; 1.6-2.6 mm 2 as average and more than 2.6 mm 2 as large. Mean and 95% confidence limits of the disc diameter by 90D examination- both with and without magnification factor correction- were determined in the corresponding groups.


  Results Top


The characteristic features of the 75 subjects are shown in [Table 1]. All the subjects were of Indian origin. Break-up according to the diagnosis of these 75 subjects is shown in [Table 2]. ICC for intra-observer agreement for horizontal disc diameter was 0.971 and for the vertical disc diameter was 0.937. The ICC for inter-observer agreement for horizontal disc diameter was 0.889 and for vertical disc diameter was 0.912. ICC for agreement between clinical and HRT horizontal disc diameter was 0.518 and for vertical disc diameter was 0.487.

The agreement between the clinical and HRT measurements of disc diameter was analyzed by the Bland and Altman method [Figure 1] and [Figure 2]. The mean difference was 0.17 (95% CI, 0.13- 0.47) for horizontal disc diameter and 0.22 (95% CI, 0.11- 0.54) for vertical diameter.

The mean (95% CI) vertical disc diameter of the three groups based on the disc size is shown in [Table 3]. It was 1.55 (1.2-1.9) in the small disc group, 1.91 (1.87-1.96) in the average disc group, and 2.15 (2.03-2.27) in the large disc group. When the direct measurement from the slit lamp was analyzed without correction for the magnification factor, the mean (95% CI) vertical disc diameter was 1.1 (0.85, 1.34) in the small disc group, 1.36 (1.32, 1.39) in the average disc group and 1.52 (1.44, 1.61) in the large disc group.


  Discussion Top


It has been shown by many studies that measurements of ONH parameters are confounded by the optic disc size. [1],[2],[3] To address this issue in a clinical perspective, it is not necessary to accurately measure the disc size but it is enough to categorize discs into small, average and large discs. [8] HRT measurements of the optic disc size are well documented for different ethnic groups. Since the access to such costly instruments is restricted; it is useful to have a clinical means of categorizing the disc by its size. Measuring the disc diameter on slit lamp biomicroscopy with a high-powered convex lens is a good alternative. In this study we report the validation and potential use of this method.

Good intra and inter observer agreement are prime requirements for any diagnostic test. The ICC for intra and interobserver agreement for vertical disc diameter measurement was 0.937 and 0.912 respectively. This is in agreement with the intra- and interobserver variability of optic measurements by slit lamp biomicroscopy. [10],[11] The other advantages of this technique for determining the optic disc size is easy accessibility compared to other expensive imaging techniques.

HRT was chosen for comparison in this study as the optic disc measurement with HRT correlated well with the measurements from histological studies, [12] in vivo measurements of optic disc diameter performed during vitrectomy [13] and also the good reproducibility reported with HRT. [14]

There have been studies looking at the agreement between fundoscopy and HRT for cup-disc ratio, [15],[16] but those for the disc diameter are limited.

Lim et al . compared the vertical disc diameters obtained with 60, 78 and 90 D lenses with that of HRT. [17] They found that the correlation was moderate to good between the two (r=0.80 with 60D lens and r=0.59 with 90D lens). In their study, optic disc size measured with HRT was larger than that with 90D lens. In our study, disc diameter measurement with HRT was less than that obtained with clinical method. Instead of correlation, we calculated ICC because correlation gives the strength of relationship between the methods while ICC gives the actual agreement between the methods. [18]

Barkana et al . [19] assessed the interchangebility of optic disc diameter measurements with slit-lamp biomicroscopy with a 78D lens and HRT 2. The mean difference between HRT and 78D lens measurements was 0.12 mm and the limits of agreement between these methods were poor (-0.29 to 0.53 for HRT and slit-lamp microscopy). In our study, the mean difference between clinical and HRT measurements as analysed on Bland and Altman plots was 0.17 (95% CI, 0.13- 0.47) for horizontal disc diameter and 0.22 (95% CI, 0.11- 0.54) for vertical diameter. In Barkana, et al .'s study too, disc diameter measurement with HRT was greater than the clinical measurement. Barkana et al . also evaluated the agreement between these methods for classification of discs into small, average and large disc categories and this was again less than moderate (Kappa statistics κ = 0.154, 95% CI 0-0.39). They concluded that HRT and clinical disc measurement methods cannot be used interchangeably in clinical practice and research.

Disc diameter measurement with HRT was always lesser than that by clinical method in our study which is contrary to that found by the above two studies. We cannot explain this but it might be due to the operator dependent subjectivity involved in delineating the disc margin in HRT.

Jonas recommended that in clinical practice, it is enough to document the disc size as average, small or large. [8] There is no validated method to classify discs into small, average and large categories. We have used the classification proposed by manufacturers of HRT. [9] These measurements would help the clinician to classify optic nerve heads into small, medium and large on routine clinical evaluation without dependence on expensive technology. It is ideal to have the cup-disc ratio corrected for the disc size but short of that, the above classification can help one corroborate with other clinical signs and decide if a 0.3:1 cup is pathological in a small disc or if a 0.7:1 cup is physiological in a large disc. We tried to group discs into these categories based on the HRT data and tried to provide the corresponding clinical disc diameter measurements, which would benefit those not using HRT. We have also provided the direct measurements from the slit lamp without magnification correction for the lens so that those using a 90 D lens can directly classify the discs into small, average or large without performing any complex calculations.

One of the limitations of the study is the use of a non-consecutive sample. As the study was basically about determining the agreement between two methods and did not involve summarizing disease characteristics, a non-consecutive sample would not have affected the results to a great extent.

This study is also limited by the number of small and large discs, which is reflected in the overlapping confidence intervals of the groups. While the sample size is adequate for agreement study, it is not adequate for classification of disc size into groups. Larger sample size with tighter confidence intervals is needed for this.

In conclusion, our study is in agreement with the previous studies. The agreement between measurements by HRT and the clinical method is only moderate. Disc diameter measurement on stereobiomicroscopy can be used to categorize discs into small, average and large discs.


  Acknowledgement Top


Hyderabad Eye Research Institute.

 
  References Top

1.
Healey PR, Mitchell P, Smith W, Wang JJ. Relationship between cup-disc ration and optic disc diameter: The Blue Mountains Eye Study. Aust N Z Ophthalmol 1997;25 Suppl 1:S99-101  Back to cited text no. 1
    
2.
Jonas JB, Berqua A, Schmitz-Valckenberg P, Papastathopoulos KI, Budde WM. Ranking of optic disc variables for detection of glaucomatous optic nerve damage. Invest Ophthalmol Vis Sci 2000;41:1764-73  Back to cited text no. 2
    
3.
Hoffmann EM, Zangwill LM, Crowston JG, Weinreb RN. Optic disk size and glaucoma. Surv Ophthalmol 2007;52:32-49  Back to cited text no. 3
[PUBMED]  [FULLTEXT]  
4.
Mansour AM. Racial variation of optic disc size. Ophthalmic Res 1991;23:67-72  Back to cited text no. 4
[PUBMED]    
5.
Varma R, Tielsch JM, Quigley HA, Hilton SC, Katz J, Spaeth GL, et al . Race-, age-, gender-, and refractive error related differences in normal optic disc. Arch Ophthalmol 1994;112:1068-76  Back to cited text no. 5
[PUBMED]  [FULLTEXT]  
6.
Akman A, Oram O, Avdin P. Optic disc measurements with 78 diopter lens, Zeiss 4-mirror contact lens and computerized image analyzing system. Eur J Ophthalmol 1998;8:22-7  Back to cited text no. 6
    
7.
Sung VC, Bhan A, Vernon SA. Agreement in assessing optic discs with a digital stereoscopic optic disc camera and HRT. Br J Ophthalmol 2002;86:196-202  Back to cited text no. 7
[PUBMED]  [FULLTEXT]  
8.
Jonas JB, Papastathopoulos K. Ophthalmoscopic measurement of optic disk. Ophthalmology 1995;102:1102-1106  Back to cited text no. 8
[PUBMED]    
9.
Fingeret M. Using the Heidelberg Retina Tomograph II (HRT II): Image Acquisition and Accessing the Data. In: Fingeret M, Flanagan JG, Liebmann JM, editors. The Essential HRT Primer. San Ramon: Jocoto Advertising, Inc.; 2005. page 25  Back to cited text no. 9
    
10.
Haslett RS, Batterbury M, Cuypers M, Cooper RL. Inter-observer agreement in clinical optic disc measurement using a modified 60d lens. Eye 1997;11(Pt 5):692-7  Back to cited text no. 10
[PUBMED]    
11.
Spencer AF, Vernon SA. Optic disc measurement: A comparison of indirect ophthalmoscopic methods. Br J Ophthalmol 1995;79:910-5  Back to cited text no. 11
[PUBMED]  [FULLTEXT]  
12.
Quigley HA, Brown AE, Morrison JD, Drance SM. The size and shape of the optic disc in normal human eyes. Arch Ophthalmol 1990;108:51-7  Back to cited text no. 12
[PUBMED]  [FULLTEXT]  
13.
Bartz-Schmidt KU, Weber J, Heimann K. Validity of two dimensional data obtained with the HRT as verified by direct measurements in normal optic nerve heads. Ger J Ophthalmol 1994;3:400-5  Back to cited text no. 13
[PUBMED]    
14.
Rohrschneider K, Burk Ro, Kruse FE, Volcher HE. Reproducibility of the optic nerve head topography with a new laser tomographic scanning device. Ophthalmology 1994;101:1044-9  Back to cited text no. 14
    
15.
Jayasundera T, Danesh-Meyer HV, Donaldson M, Gamble G. Agreement between stereoscopic photographs, clinical assessment, Heidelberg retina tomograph and digital stereoscopic optic disc camera in estimating vertical cup:disc ratio. Clin Experiment Ophthalmol 2005;33:259-63  Back to cited text no. 15
[PUBMED]  [FULLTEXT]  
16.
Watkins R, Panchal L, Uddin J, Gunvant P. Vertical cup-to-disc ratio: agreement between direct ophthalmoscopic estimation, fundus biomicroscopic estimation, and scanning laser ophthalmoscopic measurement. Optom Vis Sci 2003;80:454-9  Back to cited text no. 16
[PUBMED]  [FULLTEXT]  
17.
Lim CS, O'Brien C, Bolton NM. A simple clinical method to measure the optic disk size in glaucoma. J Glaucoma 1996;5:241- 245  Back to cited text no. 17
[PUBMED]    
18.
Sackett DL, Haynes RB, Guyatt GH, Tugwell P. Clinical Epidemiology - A Basic Science for Clinical Medicine. 2 nd Edition. Philadelphia, USA: LippinCott Williams and Wilkins;1991:55-56  Back to cited text no. 18
    
19.
Barkana Y, Harizman N, Gebber Y, Liebmann JM, Ritch R. Measurement of optic disk size with HRT II, Stratus OCT and fundoscopy are not interchangeable. Am J Ophthalmol 2006;142:375-380  Back to cited text no. 19
    


    Figures

  [Figure 1], [Figure 2]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]


This article has been cited by
1 Focused analysis of RNFL decay in glaucomatous eyes using circular statistics on high-resolution OCT data
Md. Hasnat Ali, Meghana Ray, S. Rao Jammalamadaka, Sirisha Senthil, M. B. Srinivas, Saumyadipta Pyne, Natasha Gautam
PLOS ONE. 2023; 18(10): e0292915
[Pubmed] | [DOI]
2 An innovative and simple method, clinically comparable to high-definition optical coherence tomography in quantifying posterior segment lesions in the retina
ShajiP Koshy, LeenaMariyam Varghese, Satish Thomas, Pramod Thomas
Indian Journal of Ophthalmology. 2020; 68(5): 875
[Pubmed] | [DOI]
3 Secondary macular holes: when to jump in and when to stay out
Dhananjay Shukla
Expert Review of Ophthalmology. 2013; 8(5): 437
[Pubmed] | [DOI]
4 Agreement Among 3 Methods of Optic Disc Diameter Measurement :
Harsha L. Rao, Narendra K. Puttaiah, Jonnadula G. Babu, Rajat Maheshwari, Sirisha Senthil, Chandra S. Garudadri
Journal of Glaucoma. 2010; 19(9): 650
[VIEW] | [DOI]



 

Top
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
Abstract
Materials and Me...
Results
Discussion
Acknowledgement
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed4715    
    Printed146    
    Emailed4    
    PDF Downloaded675    
    Comments [Add]    
    Cited by others 4    

Recommend this journal