Evaluation of intrastromal corneal ring segments for treatment of post-LASIK ectasia patients with a mechanical implantation technique


Department of Ophthalmology, Maltepe University School of Medicine, Istanbul, Turkey

Date of Submission02-Jan-2011
Date of Acceptance06-Jun-2011
Date of Web Publication19-Oct-2011

Correspondence Address:
Zeki Tunc
Department of Ophthalmology, Kultur mah Kucuk Camlik Sit A1 Blok D3 Etiler, 34340 Istanbul
Turkey
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0301-4738.86310

Rights and Permissions
  Abstract 

Aim: To evaluate the clinical outcomes of Keraring segment implantation in patients with post- laser-assisted in situ keratomileusis (LASIK) ectasia, using a mechanical implantation technique. Materials and Methods: Twelve eyes of 10 patients with post-LASIK ectasia were enrolled. Intracorneal ring segments (ICRS) were implanted after dissection of the tunnel using Tunc's specially designed dissector under suction. A complete ophthalmic examination was performed, including uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), spherical equivalent, keratometric (K) readings, inferosuperior asymmetry index (ISAI), and ultrasound pachymetry. All 3, 6, and 12-month follow-ups were completed, and statistical analysis was performed. Results: The mean preoperative UDVA for all eyes was 1.28 ± 0.59 logMAR. At 12 months, the mean UDVA was 0.36 ± 0.19 logarithm of the Minimum Angle of Resolution (logMAR) (P=0.002), and the mean preoperative CDVA was 0.58 ± 0.3 logMAR, which improved to 0.15 ± 0.12 (P=0.002) at 1 year. There was a significant reduction in cylindrical refractive and spherical equivalent refractive error from -5.29 ± 2.47 diopters (D) and -5.54 ± 5.04 D preoperatively to -1.47 ± 0.71 D and -0.74 ± 1.07 D (P=0.001, P=0.002), respectively, at 1 year. In the same period, the mean K- readings improved from 47.93 ± 4.84 D to 40.87 ± 2.36 D (P=0.002), and the mean ISAI improved from 5.34 ± 3.05 to 2.37 ± 1.68 (P=0.003). No significant changes in mean central corneal thickness were observed postoperatively. There were no major complications during or after surgery. Conclusion: ICRS implantation using a unique mechanical dissection technique is a safe and effective treatment for post-LASIK ectasia. All parameters showed improvement at 1-year follow-up.

Keywords: Intrastromal corneal ring segments; keraring; post-LASIK ectasia


How to cite this article:
Tunc Z, Helvacioglu F, Sencan S. Evaluation of intrastromal corneal ring segments for treatment of post-LASIK ectasia patients with a mechanical implantation technique. Indian J Ophthalmol 2011;59:437-43

How to cite this URL:
Tunc Z, Helvacioglu F, Sencan S. Evaluation of intrastromal corneal ring segments for treatment of post-LASIK ectasia patients with a mechanical implantation technique. Indian J Ophthalmol [serial online] 2011 [cited 2024 Mar 28];59:437-43. Available from: https://journals.lww.com/ijo/pages/default.aspx/text.asp?2011/59/6/437/86310

Corneal ectasia after laser-assisted in situ keratomileusis (LASIK) surgery is a rare but devastating complication that was first described by Seiler et al, [1] in 1998. Keratectasia is defined as a progressive steepening of the corneal curvature, with or without associated central and paracentral corneal thinning. Clinically, it presents as the initial appearance of low myopia that progresses over time to high myopia and irregular astigmatism, resulting in loss of uncorrected distance visual acuity (UDVA) and corrected distance visual acuity (CDVA). [2] Two major risk factors are thought to be responsible for this complication: operating on corneas with preexisting disposition to corneal ectasia, and removing too much corneal tissue. [3] After LASIK, the cornea is structurally weakened not only by the laser central stromal ablation (depending on the attempted correction), but also by the creation of the flap itself. [4] Despite the availability of this information, ectasia remains unpredictable. [5]

This ectatic disorder has an estimated incidence that ranges from 0.04% to 0.6%. [6] Treatment options for postoperative LASIK keratectasia include rigid contact lenses and lamellar or penetrating keratoplasty. Recently, intrastromal corneal ring segments (ICRS) and corneal collagen cross-linking have added a new dimension to the management of keratectasia. Furthermore, long-term data on ICRS procedures indicate the possibility of deferring or even replacing keratoplasty in keratectasia patients. [7]

Corneal tunnelization for ring segment insertion can be performed by mechanical dissection or by femtosecond laser technology. For mechanical dissection, there is already a 6 and 7 mm semiautomated dissector that operates under suction (Intacs, Addition Technology). However, in this study, 5 mm optical zone rings (Keraring) were implanted after dissection of the tunnel by using a semiautomated dissector operating under suction [Figure 1], which was designed by Dr. Tunc, in patients with post-LASIK ectasia.
Figure 1: Intrastromal dissector for a 5.0 mm diameter implantation zone with suction ring

Click here to view



  Materials and Methods Top


This prospective, non-comparative study was approved by the Ethics Board Committee and followed the tenets of the Declaration of Helsinki. All patients agreed to participate in the study and to return for the postoperative examinations. Written consent was obtained from every patient after the purpose and procedures of the study had been fully explained. Inclusion criteria were clear central corneas and no visual dysfunctions other than post-LASIK ectasia. A corneal thickness of 400 μm at the site of segment implantation was considered the minimum acceptable for the study. Exclusion criteria were additional severe ocular pathologies (e.g., glaucoma, cataract, diabetic retinopathy, and age-related macular degeneration). All operations were performed by the same surgeon (ZT) at the Department of Ophthalmology, Maltepe University School of Medicine, Istanbul, Turkey

During each surgery session, only one eye of each patient was implanted with the Keraring; if patients needed ring segment implantations for both eyes, each eye was implanted during a different session. A complete ophthalmic examination was performed preoperatively and postoperatively, including UDVA, CDVA, spherical equivalent (SE), keratometric (K) readings, the inferosuperior asymmetry index (ISAI), and ultrasound pachymetry. ISAI was calculated as the difference between the dioptric powers at 3 mm, of the corneal geometric center using EyeSys Vision, Inc V 4.5. Posterior ectasia and corneal thickness were measured using the Orbscan II Slit Scanning Corneal Topography/Pachymetry System (version 3.10.27, Orbtek Inc.) and the DGH-500 ultrasound pachymeter (DGH, Paghette 2). ICRS placement was analyzed by Fourier-domain optical coherence tomography (OCT) (Optovue RTVue with Cornea/Anterior Module (CAM)). Visual acuity was measured using Snellen notation and then converted to logMAR for statistical analysis.

The diagnosis of corneal ectasia was based on the following criteria: (1) various degrees of progressive significant myopic regression after LASIK; (2) loss of one or more Snellen lines of CDVA after LASIK; (3) ISAI more than 1.2; (4) a maximum posterior surface elevation from the best-fit sphere of 0.071 mm; (5) a maximum anterior surface elevation from the best-fit sphere of 0.042 mm; and (6) progressive corneal thinning or pachymetry less than 400 μ. [7],[8] At least 4 of the above mentioned criteria were required to be present for the diagnosis of ectasia.

Twelve eyes of 10 patients with post-LASIK ectasia underwent 5 mm intrastromal corneal ring (Keraring, Mediphacos, Belo Horizonte, Brazil) implantation [Figure 2]A and B. The mean age of patients was 25.90 ± 7.89 years, and there were 6 female and 4 male patients. The mean time to operation after LASIK was 5.6 ± 1.64 years. Preoperatively, the location of the iatrogenic cone was central in 5 eyes (41.6%), inferior in 2 eyes (16.8%), and inferotemporal in 5 eyes (41.6%). The follow-up period for all patients was at least 12 months. Corneal tunnels were made manually by using a special dissector under suction that was designed by Dr. Tunc. The technique used to place the Keraring segments in these post-LASIK corneas was identical to that used in the unoperated eyes. Because the typical flap diameter is 8.5 to 9.0 mm, and Keraring inserts had an inner diameter of 4.7 mm and an outer diameter of 5.4 mm, the inserts were positioned away from the flap border [Figure 2]B and C.
Figure 2: (A) A slit lamp photograph of a post-LASIK eye with ectasia after implantation of a new 210° Keraring segment. (B,C) Adequate distance between the Keraring segment and the post-LASIK fl ap border.

Click here to view


All surgical procedures were performed under topical anesthesia. Proparacaine hydrochloride 0.5% (Alcaine, Alcon) drops were used for topical anesthesia. The operation microscope (Carl Zeiss Meditec, Jena, Germany) was used to mark the Purkinje reflection as the central point of intrastromal corneal ring implantation. After marking the center, intraoperative ultrasonic pachymetry was performed at the site of the future incision. The diamond blade was set at 80% of the measured corneal thickness and was used to create a single radial incision of 1.2 mm at the steepest meridian [Figure 3]A and B. The mean depth of implantation was 407.58 ± 23.28 mm (range 375 mm to 442 mm). A surgical pocket for each eye was created using a micro-dissector and Suarez spreader. The intrastromal dissection was created to the full depth of the incision, and the pocket was extended along its full length. All eyes were fixed using a suction ring that was placed around the limbus and was guided by the previously marked geometric center of the cornea. The tunnels were created by using left and right Dr. Tunc's special dissectors. As a result, two 180-degree or one 240-degree semicircular dissections into the stroma with an approximate diameter of 5 mm were achieved. After the suction device, which was the same device used in Intacs segment implantation, was removed, the 1 or 2 Keraring segments were inserted into each of the semicircular channels using Albertazzi forceps [Table 1]. The 5-mm diameter Keraring with appropriate arc length and thickness was implanted in each eye according to the manufacturer's nomogram [Figure 4]. The decision to perform asymmetrical implantation was made according to the corneal topography. [9] The radial incision was closed with 1 embedded 10-0 nylon suture. All operations were uneventful. The corneal suture was removed 1 week after surgery to minimize the potential for induced astigmatism, and to minimize the risk of keratitis. Postoperative medications included a topical tobramycin 0.3% (Tobrex, Alcon), fluorometholone 5% (Flarex, Alcon), and hydroxypropyl methylcellulose 0.003% (Tears Naturale Free, Alcon) 4 times a day for 2 weeks.
Figure 3: (A) Fourier-domain optical coherence tomography (OCT) showing the precise cross-sectional visualization and ICRS at the adequate depth of the cornea and (B) the pachymetry map

Click here to view
Figure 4: Keraring segment implantation nomogram, which describes the implantation of one or two segments according to the localization of the ectasia in the corneal topography map

Click here to view
Table 1: Data of implanted rings and preoperative corneal status

Click here to view


Statistical Method

Statistical calculations were performed with the Number Cruncher Statistical System (NCSS) 2007 program for Windows. In addition to the standard descriptive statistical calculations [mean, standard deviation, and median inter-quartile range (IQR)], Friedman's repeated-measures test was used to determine the differences in measurement at each time point, and Dunn's multiple comparisons post-hoc tests were used for pair wise comparisons. The statistical significance level was established at P<0.05.


  Results Top


Twelve eyes of 10 patients underwent Keraring implantation. There were no intraoperative complications, such as anterior chamber perforation; postoperatively, all eyes showed excellent corneal tolerance with intrastromal segments. Some of the eyes showed subconjunctival hemorrhage due to the suction ring of the dissector, which dissolved spontaneously and completely after 1 week.

The mean UDVA increased significantly from 1.28 ± 0.59 logMAR preoperatively to 0.37 ± 0.2 logMAR (n=12) (P=0.002) 3 months after implantation, and at 12 months the mean UDVA was 0.36 ± 0.19 logMAR (n=12) (P=0.002) [Figure 5]A. The mean preoperative CDVA was 0.58 ± 0.3 logMAR. The mean CDVA improved to 0.16 ± 0.14 logMAR (n=12) (P=0.002) at three months after implantation. One year postoperatively, the mean CDVA was 0.15 ± 0.12 logMAR (n=12) (P=0.002) [Figure 5]B. There was a significant reduction in cylindrical refractive and spherical equivalent refractive error from -5.29 ± 2.47 diopters (D) and -5.54 ± 5.04 D preoperatively to -1.47 ± 0.71 D and -0.74 ± 1.07 D (P=0.001, P=0.002) respectively at 1 year [Table 2]. In the same period, the mean K readings improved from 47.93 ± 4.84 D to 40.87 ± 2.36 D (P=0.002) [Figure 6]A and B and the mean ISAI from 5.34 ± 3.05 to 2.37 ± 1.68 (P=0.003). No significant changes in preoperative mean central corneal thickness (439 ± 52 μm) were observed 1 year postoperatively (440 ± 50 μm). [Table 3] shows the means, standard deviation and median inter-quartile range (IQR) of all data. Significant improvements were observed postoperatively for UDVA and CDVA, and corneal ectasia, keratometry, SE, and ISAI were significantly reduced. Moreover, according to Dunn's multiple comparisons tests, 3, 6, 12-month postoperative UDVA, CDVA, keratometry, and SE values showed significantly improved results from the preoperative values. Additionally, 3, 6, and 12-month results showed that there was no significant difference in refraction stability [Table 4].
Figure 5: Graphs showing the comparison of (A) uncorrected distance visual acuity (UDVA) and (B) corrected distance visual acuity (CDVA) values between the preoperative test and the 3, 6, and 12-month postoperative follow-ups

Click here to view
Figure 6: (A,B) The topography of the eye with post-LASIK irregular astigmatism before (top left) and 12 months after (bottom left) ICRS implantation and the difference map (right)

Click here to view
Table 2: Preoperative and postoperative 3, 6, and 12-month follow-up data

Click here to view
Table 3: Statistical calculations of preoperative and postoperative 3, 6, and 12-month follow-up data with Friedman's test

Click here to view
Table 4: Evaluation of the postoperative results and the refraction stability with Dunn's multiple comparisons test

Click here to view


The patients were subdivided into two groups: Those with one (group 1, n=5) and two (group 2, n=7) segment implantations. The outcomes were further analyzed for these groups. The preoperative central pachymeter and incision depth of the groups were 461.60 ± 40.99 and 422.57 ± 55.50 μm, respectively, and 408.80 ± 23.86 and 406.71 ± 24.74 μm, respectively and statistical analysis showed no significant difference for these parameters (P1=0.214 P2=0.887). The preoperative mean UDVA of groups 1 and 2 were 1.04 ± 0.32 and 1.44 ± 0.69 logMAR, respectively which were not significantly different (P=0.262). The preoperative mean CDVA of groups 1 and 2 were 0.46 ± 0.21 and 0.67 ± 0.34 logMAR, respectively (P=0.253). The preoperative mean spherical equivalent refractive errors of the groups 1 and 2 were 2.67 ± 1.15 D and 7.94 ± 5.23 D, respectively, and group 2 had a significantly higher refractive error than group 1 (P=0.038). The preoperative mean cylindrical refractive errors of the groups 1 and 2 were -3.95 ± 1.69 D and -6.25 ± 2.59 D, respectively, and group 2 had a insignificantly higher refractive error than group 1 (P=0.116). The preoperative mean K readings of the groups were 45.50 ± 1.63 D and 49.66 ± 5.72 D, respectively. The mean K readings of group 2 were slightly but not significantly higher than the K readings of group 1, (P=0.150). The preoperative mean ISAI of the groups were 5.62 ± 2.99 and 5.13 ± 3.30 (P=0.798), respectively. Improvements in all of the parameters in both groups were observed at 12 months [Table 2] and [Table 4]. The 12 -month postoperative mean UDVA and CDVA of group 1 and 2 were 0.31 ± 0.19 and 0.38 ± 0.19 logMAR, respectively, and 0.08 ± 0.04 and 0.19 ± 0.14 logMAR, respectively (p1=0.558, p2=0.108). The 12-month postoperative mean spherical equivalent refractive errors of the groups 1 and 2 were 0.82 ± 0.73 D and 1.17 ± 0.79 D, respectively (P=0.452). The 12-month postoperative mean cylindrical refractive errors of the groups 1 and 2 were -1.15 ± 0.60 D and -1.71 ± 0.74 D, respectively (P=0.192). The 12-month postoperative mean K readings of the groups were 41.49 ± 2.01 D and 40.41 ± 2.63 D, respectively (P=0.462). The 12-month postoperative mean ISAI of the groups were 2.63 ± 1.11 and 2.17 ± 2.05, respectively (P=0.659).


  Discussion Top


Ectasia after LASIK is an uncommon, but potentially visually disabling complication. [10] Despite the number of studies that support the efficacy of LASIK, [11] concerns about the occurrence of postoperative keratectasia are growing. Even though corneal ectasia is a rather rare complication after LASIK, it can have a profoundly negative effect on the refractive properties of the cornea. One of the possible alternatives to manage post-LASIK corneal ectasia is the implantation of ICRS, which were initially developed for the correction of myopia. [12],[13] Recently, corneal collagen cross-linking with riboflavin and ultraviolet-A light with or without ICRS implantation is used to slow down the progression of ectasia. [14] ICRS were designed to achieve a refractive adjustment by flattening the central corneal curvature while maintaining clarity in the central optical zone. Because of the removable and tissue-saving nature of this technique, its application could be expanded to patients with corneal thinning disorders. Several reports have demonstrated the efficacy of Intacs in correcting keratoconic eyes, [15],[16],[17],[18] and preliminary studies show encouraging results in eyes with post-LASIK corneal ectasia. [7],[8],[19],[20],[21] The magnitude of this flattening effect is in direct proportion to the thickness of the implant and in inverse proportion to its diameter. [22] Furthermore, soft-ectasia corneas show more flattening than healthy corneas. [16] The goal of implanting intracorneal rings for keratectasia after LASIK is to reduce the corneal steepening, which results in a favorable visual outcome and eliminates or delays the need for corneal grafting.

The 3 main ICRS in the market are Intacs (Addition Technology, Inc.), Ferrara (Ferrara Ophthalmics Ltd.), and Keraring (Mediphacos Ltd.). Kerarings are made of polymethylmethacrylate (PMMA) and are characterized by a triangular cross-section that induces a flattening effect on the cornea. Their apical diameter is 5 mm, and the flat basis width is 0.6 mm, with variable thickness (0.15 to 0.35 mm with 0.05-mm steps) and arc length (90 degrees, 160 degrees, and 210 degrees). The optical zone provided by Keraring segments is 5.0 mm in diameter.

The surgical technique for tunnel creation in Keraring and Ferrara procedures differs from the mechanical tunnel creation technique used in Intacs. While a suction device is used in Intacs, the tunnels of Keraring and Ferrara are prepared by the Ferrara double spatula. Implantation of ICRS without the use of a suction device is a more surgeon dependent technique, which has increased risk of complications while the surgeon is in the learning curve of the procedure. Additionally, Kwitko and Severo reported decentration of Ferrara ICRS in 3.9% of cases, segment extrusion in 19.6%, and bacterial keratitis in 1.9%. [23] The authors suggest that most of the complications related to surgical technique were caused by the surgeon's learning curve, and the differing healing processes of keratoconic corneas. Siganos et al. reported superficial implantation and asymmetric placement of the segments in 7.7% of cases. A vacuum was not used to create the channels in their series, and these complications were associated with implantation of the Ferrara rings. [24]

Making the tunnels in the 5 mm optical zone gives the surgeon the ability to correct the higher amount of refractive error with thinner ICRS implantation. However, this procedure requires precision because the cornea in this area is thinner than the periphery of the cornea and is close to the optical axis [Figure 3]A and B. Recently, Kubaloglu et al. used suction rings (Moria) to minimize decentration during the implantation of Keraring segments. [25] Unfortunately, this suction ring only fixes the eye but cannot prevent uncontrolled movements of the dissector. Although the operations were performed by an experienced surgeon, corneal perforation and superficial segment implantation was observed in 6% of the cases who had mechanical implantation. [26] We believe that fixating the eye under a suction during intrastromal dissection is essential to be able to create a tunnel at the adequate depth preoperatively. In our cases, the intrastromal dissection was created to the full depth of the incision. The suction ring not only served to fix the eye but also prevented uncontrolled movement of the dissector during rotational movement (counterclockwise and clockwise) by holding the dissector (like the semiautomated suction ring of Intacs). The tunnels were created by using left and right special dissectors with a device containing a semiautomated suction ring. As a result, two 180-degree or one 240-degree semicircular dissections into the stroma with an approximate diameter of 5 mm were achieved at the adequate depth and with proper centralization [Figure 3]A. All operations were uneventful. There was no incidence of delayed complications. The integrity of the cornea was well preserved in all eyes, and there was no extrusion of the rings. In contrast, standard mechanical stromal dissection of the Keraring could cause a higher rate of extrusion. [23],[24],[25],[26]

Alio΄ et al, found significant improvement in visual acuity after Intacs segment implantation in three eyes with ectasia. In 2 eyes, the UDVA was 20/40 postoperatively. In the third eye, there was a residual refractive error; and, the UDVA was 20/50, and the CDVA was 20/40. [7] In a post-LASIK ectasia study, Kymionis et al, implanted Intacs segments in eyes with a mean preoperative UDVA of 20/100. At the last follow-up examination, 6 (75%) of 8 eyes had a UDVA 20/40 or better. At the end of the first postoperative year, UDVA, CDVA, and topography were stable and remained so during the follow-up period. [19] In a recent study of single Ferrara segment implantation in patients with corneal ectasia after refractive surgery by Torquetti et al., both UDVA and CDVA increased significantly after surgery (P=0.005, P=0.008). Manual implantation technique was used in their surgeries and very experienced surgeons performed the operations. No complications was seen in their series of 25 eyes. [27] Our study has potential limitations, such as the small sample of treated eyes, the lack of higher-order aberration analysis and the lack of a comparative group. However, the results are similar to those in a post-LASIK study in which ICRS were used for treatment. [27]

Our postoperative results reveal a significant reduction in the magnitude of corneal steepening, an increase in topographical regularity, and an improvement in the UDVA due to the improved ISAI and simultaneous partial correction of the spherical and cylindrical refractive errors. Tunc's dissector for the preparation of the tunnel facilitates the procedure and adds to the safety of the surgery in patients with post-LASIK ectasia. Further follow-up and additional cases are needed to draw final conclusions about the efficacy of this surgical technique.

 
  References Top

1.
Seiler T, Koufala K, Richter G. Iatrogenic keratectasia after laser in situ keratomileusis. J Refract Surg 1998;14:312-7.  Back to cited text no. 1
    
2.
Randleman JB, Russell B, Ward MA, Thompson KP, Stulting RD. Risk factors and prognosis for corneal ectasia after LASIK. Ophthalmology 2003;110:267-75.  Back to cited text no. 2
    
3.
Flanagan GW, Binder PS. Estimating residual stromal thickness before and after laser in situ keratomileusis. J Cataract Refract Surg 2003;29:1674-83.  Back to cited text no. 3
    
4.
Kymionis GD, Tsiklis NS, Pallikaris IG, Kounis G, Diakonis VF, Astyrakakis N, et al. Long-term Follow-up of Intacs for Post-LASIK Corneal Ectasia. Ophthalmology 2006;113:1909-17.  Back to cited text no. 4
    
5.
Rodríguez LA, Guillén PB, Benavides MA, Garcia L, Porras D, Daqui-Garay RM. Penetrating keratoplasty versus intrastromal corneal ring segments to correct bilateral corneal ectasia: Preliminary study. J Cataract Refract Surg 2007;33:488-96.  Back to cited text no. 5
    
6.
Randleman JB, Stulting RD. Corneal ectasia. In: Alió JL, Azar DT, editors. Management of Complications in Refractive Surgery. Berlin: Springer Inc.; 2008. p. 89-96.  Back to cited text no. 6
    
7.
Alió J, Salem T, Artola A, Osman A. Intracorneal rings to correct corneal ectasia after laser in situ keratomileusis. J Cataract Refract Surg 2002;28:1568-74.  Back to cited text no. 7
    
8.
Twa MD, Nichols JJ, Joslin CE, Kollbaum PS, Edrington TB, Bullimore MA, et al.. Characteristics of Corneal Ectasia After LASIK for Myopia. Cornea 2004;23:447-57.  Back to cited text no. 8
    
9.
Alio´ JL, Shabayek MH, Belda JI, Correas P, Feijoo E. Analysis of results related to good and bad outcome of Intacs implantation for correction of keratoconus. J Cataract Refract Surg 2006;32:756-61.  Back to cited text no. 9
    
10.
Binder PS. Ectasia after laser in situ keratomileusis. J Cataract Refract Surg 2003;29:2419-29.  Back to cited text no. 10
    
11.
Sugar A, Rapuano CJ, Culbertson WW, Huang D, Varley GA, Agapitos PJ, et al. Laser in situ keratomileusis for myopia and astigmatism: Safety and efficacy: A report by the American Academy of Ophthalmology. Ophthalmology 2002;109:175-87.  Back to cited text no. 11
    
12.
Burris TE. Intrastromal corneal ring technology: Results and indications. Curr Opin Ophthalmol 1998;9:9-14.  Back to cited text no. 12
    
13.
Nosé W, Neves RA, Burris TE, Schanzlin DJ, Belfort Júnior R. et al. Intrastromal corneal ring: 12-month sighted myopic eyes. J Refract Surg 1996;12:20-8.  Back to cited text no. 13
    
14.
Agrawal WB. Corneal collagen cross-linking with riboflavin and ultraviolet-A light for keratoconus: Results in Indian eyes. Indian J Ophthalmol 2009;57:111-4.  Back to cited text no. 14
[PUBMED]  Medknow Journal  
15.
Colin J, Cochener B, Savary G, Malet F, Holmes-Higgin D. INTACS inserts for treating keratoconus: One-year results. Ophthalmology 2001;108:1409-14.  Back to cited text no. 15
    
16.
Tunc Z, Deveci N, Sener B, Bahcecioglu H. Corneal ring segments (INTACS) for the treatment of asymmetrical astigmatism of the keratoconus. Follow-up after 2 years. J Fr Ophtalmol 2003;26:824-30.  Back to cited text no. 16
    
17.
Siganos CS, Kymionis GD, Kartakis N, Theodorakis MA, Astyrakakis N, Pallikaris IG. Management of keratoconus with INTACS. Am J Ophthalmol 2003;135:64-70.  Back to cited text no. 17
    
18.
Boxer Wachler BS, Christie JP, Chandra NS, Chou B, Korn T, Nepomuceno R. INTACS for keratoconus. Ophthalmology 2003;110:1031-40.  Back to cited text no. 18
    
19.
Kymionis GD, Siganos CS, Kounis G, Astyrakakis N, Kalyvianaki MI, Pallikaris IG. Management of post-LASIK corneal ectasia with INTACS inserts: One-year results. Arch Ophthalmol 2003;12:322-6.  Back to cited text no. 19
    
20.
Pokroy R, Levinger S, Hirsh A. Single INTACS segment for post-laser in situ keratomileusis keratectasia. J Cataract Refract Surg 2004;30:1685-95.  Back to cited text no. 20
    
21.
Siganos CS, Kymionis GD, Astyrakakis N, Pallikaris IG. Management of corneal ectasia after laser in situ keratomileusis with INTACS. J Refract Surg 2002;18:43-6.  Back to cited text no. 21
    
22.
Patel S, Marshall J, Fitzke FW. Model for deriving the optical performance of the myopic eye corrected with an intracorneal ring. J Refract Surg 1995;11:248-52.  Back to cited text no. 22
    
23.
Kwitko S, Severo NS. Ferrara intracorneal ring segments for keratoconus. J Cataract Refract Surg 2004;30:812-20.  Back to cited text no. 23
    
24.
Siganos D, Ferrara P, Chatzinikolas K, Bessis N, Papastergiou G. Ferrara intrastromal corneal rings for the correction of keratoconus. J Cataract Refract Surg 2002;28:1947-51.  Back to cited text no. 24
    
25.
Kubaloglu A, Cinar Y, Sogutlu Sari E, Koytak A, Ozdemir B, Ozertürk Y, et al. Comparison of 2 intrastromal corneal ring segment models in the management of keratoconus. J Cataract Refract Surg 2010;36:978-85.  Back to cited text no. 25
    
26.
Kubaloglu A, Sogutlu Sari E, Cinar Y, Cingu K, Koytak A, Coskun E, et al. Comparison of mechanical and femtosecond laser tunnel creation for intrastromal corneal ring segment implantation in keratoconus. J Cataract Refract Surg 2010;36:1556-61.  Back to cited text no. 26
    
27.
Torquetti L, Ferrara P. Intrastromal corneal ring segment implantation for ectasia after refractive surgery. J Cataract Refract Surg 2010;36:986-90.  Back to cited text no. 27
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]


This article has been cited by
1 Long-term Follow-up of a 150° arc-length Intrastromal Corneal Ring Segment Implantation in post-LASIK ectasia.
Lucía Ibares-Frías, Luis Fernández-Vega-Cueto, Belén Alfonso-Bartolozzi, Aranzazu Poo-López, David Madrid-Costa, José F. Alfonso
Current Eye Research. 2022; : 1
[Pubmed] | [DOI]
2 Ectasia After Corneal Refractive Surgery: A Systematic Review
Majid Moshirfar, Alyson N. Tukan, Nour Bundogji, Harry Y. Liu, Shannon E. McCabe, Yasmyne C. Ronquillo, Phillip C. Hoopes
Ophthalmology and Therapy. 2021; 10(4): 753
[Pubmed] | [DOI]
3 Post Laser-assisted in situ Keratomileusis Ectasia: A Systematic Review
Adel Barbara, Yumna Busool
International Journal of Keratoconus and Ectatic Corneal Diseases. 2017; 6(2): 78
[Pubmed] | [DOI]
4 Late-onset Post-lasik Ectasia with no Apparent Risk Factor except Eye Rubbing: A Case Report and Literature Review
Ramez Barbara, Adel Barbara
International Journal of Keratoconus and Ectatic Corneal Diseases. 2016; 5(2): 85
[Pubmed] | [DOI]
5 Ferrara Intrastromal Corneal Ring Segments
Leonardo Torquetti, Jordana Sandes
International Journal of Keratoconus and Ectatic Corneal Diseases. 2016; 5(3): 114
[Pubmed] | [DOI]
6 Reshaping procedures for the surgical management of corneal ectasia
Mohammed Ziaei,Allon Barsam,Neda Shamie,David Vroman,Terry Kim,Eric D. Donnenfeld,Edward J. Holland,John Kanellopoulos,Francis S. Mah,J. Bradley Randleman,Sheraz Daya,Jose Güell
Journal of Cataract & Refractive Surgery. 2015; 41(4): 842
[Pubmed] | [DOI]
7 Femtosecond Laser-Assisted MyoRing Implantation in Postoperative LASIK Ectasia
Mahmoud Jabbarvand, Mohammadnaser Hashemian, Hesam Hashemian, Fatemeh Bazvand, Mehdi Khodaparast
Journal of Refractive Surgery. 2014; 30(7): 462
[Pubmed] | [DOI]
8 Ultrastructural and three-dimensional study of post-LASIK ectasia cornea
Saeed Akhtar,Hind Alkatan,Omar Kirat,Turki Almubrad
Microscopy Research and Technique. 2013; : n/a
[Pubmed] | [DOI]
9 Intrakorneale Ringsegmente beim Keratokonus
M. El-Husseiny,T. Tsintarakis,T. Eppig,A. Langenbucher,B. Seitz
Der Ophthalmologe. 2013; 110(9): 823
[Pubmed] | [DOI]
10 Late Onset Post-Lasik Ectasia
Lamis Abdelaziz, Joseph Pikkel, Hanna Garzozi, Ankur Barua
International Journal of Keratoconus and Ectatic Corneal Diseases. 2012; 1(3): 190
[Pubmed] | [DOI]
11 Intraoperative and postoperative complications of laserin situkeratomileusis
Jennifer G Bromley,Trent D Albright,Bhairavi Kharod-Dholakia,Joung Y Kim
Expert Review of Ophthalmology. 2012; 7(1): 25
[Pubmed] | [DOI]
12 Intraoperative and postoperative complications of laser in situ keratomileusis
Bromley, J.G., Albright, T.D., Kharod-Dholakia, B., Kim, J.Y.
Expert Review of Ophthalmology. 2012; 7(1): 25-31
[Pubmed]



 

Top
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
Abstract
Materials and Me...
Results
Discussion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed4281    
    Printed134    
    Emailed3    
    PDF Downloaded303    
    Comments [Add]    
    Cited by others 12    

Recommend this journal