ORIGINAL ARTICLE
Year : 2019  |  Volume : 67  |  Issue : 10  |  Page : 1575-1578

Central corneal thickness and corneal diameter in preterm and term newborns and preterm neonates at term


1 Department of Ophthalmology, Lady Hardinge Medical College, New Delhi, India
2 Department of Paediatrics, Kalawati Saran Children's Hospital, New Delhi, India

Date of Submission02-Mar-2019
Date of Acceptance02-May-2019
Date of Web Publication23-Sep-2019

Correspondence Address:
Dr. Anurag Shandil
Department of Ophthalmology, Lady Hardinge Medical College, New Delhi
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijo.IJO_1988_18

Rights and Permissions
  Abstract 


Purpose: To record central corneal thickness and corneal diameter in preterm and term newborns within first week of their birth and in preterm neonates at term age. Methods: Babies born at ≤34 weeks of gestation period (n = 100), term babies who have a gestation period of >37 weeks (n = 100) and preterm neonates at term age (38-42 weeks) who were born at ≤34 weeks (n = 100) were included in this hospital based observational descriptive study. Corneal diameter was taken by Castroviejo Calipers. Central corneal thickness was measured using portable ultrasonic pachymeter (Pachette 3). Refractive status of the eyes was measured using streak Wellch Allyn retinoscope under cyclopentolate cycloplegia. Results: Measure of central corneal thickness and corneal diameter in preterm newborns was found to be 633.5 ± 2.8 μm and 8.1 ± 0.6 mm, in full term newborns it was 555.1 ± 2.7 μm and 9.5 ± 0.6 mm, and in preterm neonates at term age, it was 563.5 ± 2.5 μm and 9.6 ± 0.5 mm, respectively. Mean central corneal thickness was found to be more in preterm newborns 633.5 um then term newborns 555.1 um and the difference was statistically significant (P =0.001). Mean corneal diameter was found to be maximum (9.6 mm) in preterm neonates at term age. Conclusion: Preterm babies have thick corneas and small corneal diameters. This study is useful with respect to the analysis of glaucoma in children and using preterm corneas as donor corneas in penetrating keratoplasty.

Keywords: Corneal diameter, corneal thickness, newborn


How to cite this article:
Sehrawat P, Beri S, Garg R, Datta V, Shandil A. Central corneal thickness and corneal diameter in preterm and term newborns and preterm neonates at term. Indian J Ophthalmol 2019;67:1575-8

How to cite this URL:
Sehrawat P, Beri S, Garg R, Datta V, Shandil A. Central corneal thickness and corneal diameter in preterm and term newborns and preterm neonates at term. Indian J Ophthalmol [serial online] 2019 [cited 2024 Mar 29];67:1575-8. Available from: https://journals.lww.com/ijo/pages/default.aspx/text.asp?2019/67/10/1575/267415



The cornea of preterm is different from a term newborns in parameters like corneal thickness, diameter and refractive power. Central corneal thickness (CCT) measurement plays an important role in diagnosis and treatment of various eye disorders. The number of preterm births has risen significantly in recent times with advanced medical care. Still there is very scant data on the central corneal thickness of premature and full term newborns during the neonatal period. A few studies such as Autzen and Bjørnstrøm [1] (13 premature infants), Kirwan et al.[2] (35 premature infants), and more recently by De Silva et al.[3] (56 premature infants), Uva et al.[4] (33 premature infants), Gunay et al.[5] (100 premature infants) and Jethani et al.[6] (85 premature infants) have presented their findings on the CCT in premature children.

The aim of this study is to record central corneal thickness and corneal diameter in preterm and term newborns within the first week of their birth and in preterm neonates at term age. There is very little literature about corneal thickness in preterm newborns worldwide and no study (to the best of the research knowledge) in India has been conducted for central corneal thickness and corneal diameter in preterm and term newborns and preterm neonates at term age.


  Methods Top


The current study is a Hospital based observational descriptive study which has been carried out at “Department of Ophthalmology and Department of Pediatrics and Neonatology” in a tertiary care centre from November 2013 to March 2015. The study was approved by the ethical committee of the institution.

There were 3 groups of participants. They were as follows: Group A, with 100 preterm newborns; Group B, with 100 full term newborns; Group C, with 100 preterm neonates as they reached term age. Preterm and term newborns within first week of their birth and preterm neonates at term age underwent one time examination for this study.

The non-invasive nature of the procedure was explained to the parents and appropriate consent was obtained from them. All selected preterm and term babies fulfilling the inclusion criteria were subjected to detailed birth and postnatal history. Mother's antenatal history was recorded. Those not fulfilling the inclusion criteria have been excluded.

Inclusion criteria

All the subjects with following characteristics have been included in the study:

  1. Group A: Babies born at ≤34 weeks of gestation period
  2. Group B: Term babies who have a gestation period of >37 weeks
  3. Group C: Preterm babies at term age (38-42 weeks) who were born at ≤34 weeks
  4. Clinically stable babies who are not carrying any major symptom which can make the study biased towards it
  5. Parental consent to record data.


Exclusion criteria

All the subjects with the following characteristics have been excluded from the study:

  1. Babies with any congenital ocular abnormalities
  2. Babies with congenital glaucoma
  3. Babies with congenital cataract<
  4. Babies with retinopathy of prematurity.


Eye speculum was applied and paracaine drop was instilled in the eye. The right eye was examined first and then the left. Corneal diameter was taken by Castroviejo Calipers. Central corneal thickness was measured using portable ultrasonic pachymeter. The DGH 555 Ultrasonic Pachymeter (Pachette 3) is an ultrasonic pachymeter that uses echo spike techniques to measure the thickness of the cornea. An average of three pachymetry readings were taken. The reference point here was the centre of the undilated pupil. Refractive status of the eyes was measured using streak Wellch Allyn retinoscope under cyclopentolate cycloplegia. All the measurements were recorded by the same investigator.

Finally, once the recording was done, the data sheet was prepared for the statistical software (SPSS) to run and compute various statistical tests. The results thus obtained have been presented in a graphical and tabular manner in the report.


  Results Top


The demographic profiles of the three groups were recorded.

The mean age of the preterm babies was found to be 31.7 ± 2.3 weeks while mean weight was found to be 1420 ± 600 grams. Amongst term age newborns, mean age was 38.4 ± 2.3 weeks and mean weight was 2880 ± 500 grams. The mean age of the preterm neonates at term age was found to be 39.8 ± 1.5 weeks, while mean weight was found to be 1740 ± 700 grams [Table 1].
Table 1: The post conceptional age (in weeks) and weight (in grams) distribution of the three groups

Click here to view


Mean central corneal thickness was found to be maximum 633.5 um in preterm newborns and minimum in term newborns 555.1 um and the difference was statistically significant too (P = 0.001). Mean corneal diameter was found to be maximum 9.6 mm in preterm neonates at term age and minimum in preterm newborns 8.1 mm and the difference was statistically significant too (P =0.001) [Table 2].
Table 2: Central corneal thickness and corneal diameter at birth in preterm newborn vs term newborn vs preterm neonates at term age

Click here to view


The term newborns had the highest overall refraction while the preterm neonates at term age had the lowest refraction. Amongst the 3 groups, the refraction was not very much varying for right eye, left eye and on the overall basis. Thus, it was found that refraction of preterm newborns, full term newborn, and preterm neonates at term age was found to be similar and they are hypermetropic [Table 3] and [Figure 1].
Table 3: Refractive errors at birth in preterm neonates vs term newborn vs preterm neonates at term age

Click here to view
Figure 1: (a) Distribution chart for refractive errors at birth in preterm vs term newborn vs preterm neonates at term age. (b) Distribution chart for variation of refractive errors between right and left eye at birth in preterm vs term newborn vs preterm neonates at term age

Click here to view



  Discussion Top


The preterm newborns have a thicker and smaller cornea. In our study, measure of central corneal thickness and corneal diameter in preterm newborns was found to be 633.5 ± 2.8 μm and 8.1 ± 0.6 mm respectively. Measure of central corneal thickness and corneal diameter in full term newborns was found to be 555.1 ± 2.7 μm and 9.5 ± 0.6 mm, respectively. There was no significant difference found between the right eye and left eye measurements.

These results were similar with that of the previous researchers. In Turkey, the overall mean central corneal thickness was found to be 647.7 ± 63.9 μm for babies having gestation period of 25-34 weeks i.e. preterm newborns.[7] Similarly, in England, for full term newborns, the mean central corneal thickness was measured to be 581 ± 4.7 μm.[8] Also, with respect to corneal diameter study in Saudi Arabia by KU al-Umran et al. showed that corneal diameter ranged between 7.7 mm in preterm newborns and 10 mm in full term babies.[9],[10]

In our study, the central corneal thickness and corneal diameter in preterm neonates at term age was obtained and it was found to be 563.5 ± 2.5 μm and 9.6 ± 0.5 mm respectively.

Moreover, there was no significant difference in central corneal thickness and corneal diameter of preterm neonates at term age and full term newborns. This clearly shows that preterm babies catch up with the term babies in the parameters like central corneal thickness and corneal diameter. This is similar to parameters such as weight and length of the preterm babies, which also is nearly the same in preterm babies at term age to that of the full term.

Studies in Ireland by Kirwan et al. have shown that central corneal thickness in preterm newborns decreases from 691 um at 31 weeks to 605 um at 38 weeks along with a significant increase in horizontal corneal diameter of 8.0 mm at 31 weeks to 8.9 mm at 38 weeks. These parameters further changed to central corneal thickness of 564 μm and corneal diameter 9.6 mm at 39-41 weeks age i.e., their term age.[2] The change in the 2 corneal parameters seem to be related to each other and began immediately after birth.

De Silva et al.[3] in Italy conducted a study on the thickness development in premature infants, and it was concluded that central corneal thickness decreased from 794 to 559 μm from preterm newborns at term age to term newborns. Uva et al. reported a mean CCT of 599 ± 36 μ in premature newborns and 576 ± 26 μ in full term newborns.[4] Jethani et al. did a comparison of central corneal thickness on the basis of post-gestational age which showed a mean thickness of 620.7 ± 88.8 and 574.4 ± 78.3 μ in the <260 days and >260 days age groups respectively.[6]

Our data of 633.5 ± 2.8 μm and for preterm is similar to reported by Jethani et al.,[6] but lower than the studies by Autzen and Bjørnstrøm,[1] Kirwan et al.[2] De Silva et al.[3] and Gunay et al.[5] Gunay et al. and Uva et al. also correlated birth weight and CCT. Their study on premature children suggested that the birth weight was negatively correlated with CCT.[4],[5]

Intraocular pressure measurement in glaucoma is related to the corneal thickness. Thicker corneas give a false high measurement of intra ocular pressure. Moreover, preterm newborns are associated with increased corneal hydration at birth. So, while diagnosing congenital glaucoma in a neonate, knowledge of this increased thickness would be of great significance. Although the increased corneal thickness increases the IOP recorded, there is also increased corneal hydration which may lead to ease of corneal deformation or applanation. Moses et al.[11] described underestimation of IOP in oedematous corneas due to increased applanation and corneal deformation.

Acar et al. recorded the intraocular pressure and central corneal thickness in healthy premature infants. The mean IOP was 16.8 ± 2.9 mm Hg and the mean CCT, 590 ± 58.2 μm.[12] IOP and CCT values were found to be higher in infants with lower PCA. Higher IOP values in these infants was associated with higher CCT values. Similarly, follow-up of a diagnosed glaucoma in preterm newborns would have to be done keeping in view of this change of corneal thickness up till term age.

The use of donor cornea of a preterm in corneal transplantation was always considered to be difficult. This is basically due to the increased thickness of the cornea of preterm newborns. Preterm newborns used as donor corneas in penetrating keratoplasty yield refractive errors with myopic shift because of the steep corneal configuration in preterm newborns. The surgical difficulties are encountered because of thick cornea of these preterm babies. The knowledge of this increased thickness would help in clinical implications of the use of such corneas in penetrating keratoplasty.

It was found in our study that refraction in the preterm newborns, full term newborn preterm neonates at term age was found to be similar and they are hypermetropic. However, Snir et al.[13] found that full term infants are usually hypermetropic, whereas premature babies, with or without previous retinopathy of prematurity (ROP), more often have myopia and astigmatism. The risk of myopia increases in the presence of ROP, and its magnitude increases with a decrease in the age of prematurity and an increase in severity of the residua of ROP. Babies developing ROP are considered to have anterior segment arrest. It is possible that their corneal thickness remains thicker and do not come to normal value at term age.

Limitation of our study includes small sample size and lack of follow up of preterm babies till term age as it was a one-time study. Moreover, it could have been more informative if IOP was also measured along with corneal measurements, but further studies for the same can be planned.


  Conclusion Top


This study analyzes central corneal thickness and corneal diameter amongst preterm newborns, full term newborns and preterm neonates at term age.

Major results obtained from the study suggests that central corneal thickness at birth in preterm newborn is significantly more than term newborn and preterm neonates at term age. The corneal diameter at birth in preterm newborn is significantly less than term newborn and preterm neonates at term age. The central corneal thickness and corneal diameter in preterm neonates at term age is almost equivalent to term newborn.

This study is useful with respect to the analysis of glaucoma in children and using preterm corneas as donor corneas in penetrating keratoplasty. Preterm babies have the ability to catch up with the full term in various parameters, including but not limited to corneal thickness and corneal diameter. This finding could help in computing the gestational age of preterm newborns. Also, we suggest further study of corneal thickness in babies developing retinopathy of prematurity.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Autzen T, Bjørnstrøm L. Central corneal thickness in premature babies. Acta Ophthalmol (Copenh) 1991;69:251-2.  Back to cited text no. 1
    
2.
Kirwan C, O'Keefe M, Fitzsimon S. Central corneal thickness and corneal diameter in premature infants. Acta Ophthalmol Scand 2005;83(Suppl 6):751-3.  Back to cited text no. 2
    
3.
De Silva S, Parentin F, Michieletto P, Pensiero S. Corneal curvature and thickness development in premature infants. J Pediatr Ophthalmol Strabismus 2011;48:25-9.  Back to cited text no. 3
    
4.
Uva MG, Reibaldi M, Longo A, Avitabile T, Gagliano C, Scollo D, et al. Intraocular pressure and central corneal thickness in premature and full-term newborns. J AAPOS 2011;15:367-9.  Back to cited text no. 4
    
5.
Gunay M, Celik G, Gunay BO, Dogru M, Gursoy T, Ovali HF. Central corneal thickness measurements in premature infants. Int J Ophthalmol 2014;7:496-500.  Back to cited text no. 5
    
6.
Jethani J, Shah K, Jethani M. Evaluating the change in central corneal thickness in neonates (term and preterm) in Indian population and the factors affecting it. Indian J Ophthalmol 2015;63:501-3.  Back to cited text no. 6
[PUBMED]  [Full text]  
7.
Murat G, Celik G, Gunay BO, Dogru M, Gursoy T, Ovali HF. Central corneal thickness measurements in premature infants. International J Ophthalmol 2014;7(Suppl 3):496-500.  Back to cited text no. 7
    
8.
Pennie FC, Wood ICJ, Olsen C, White S, Charman WN. A longitudinal study of the biometric and refractive changes in full-term infants during the first year of life. Vision Res 2001;41(Suppl 21):2799-810.  Back to cited text no. 8
    
9.
Al-Umran, Khaled U, Maurizio FP. Corneal diameter in premature infants. Br J Ophthalmol 1992;76(Suppl 5):292-3.  Back to cited text no. 9
    
10.
Rushood AA, Zahrani MH, Khamis A, Rushood AA. Central corneal thickness in full-term Saudi newborns. Acta Ophthalmol 2012;90(Suppl 5):355-8.  Back to cited text no. 10
    
11.
Moses RA. The Goldmann applanation tonometer. Am J Ophthalmol 1958;46:865-9.  Back to cited text no. 11
    
12.
Acar DE, Acar U, Tunay ZO, Ozdemir O, Dolgun A, Erdurmus M. The intraocular pressure and central corneal thickness in healthy premature infants. JAAPOS 2015;19:108-11.  Back to cited text no. 12
    
13.
Snir M, Friling R, Weinberger D, Sherf I, Axer-Siegel R. Refraction and keratometry in 40 week old premature (corrected age) and term infants. Br J Ophthalmol 2004;88(Suppl 7):900-4.  Back to cited text no. 13
    


    Figures

  [Figure 1]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]


This article has been cited by
1 Visco-Circumferential-Suture-Trabeculotomy Versus Rigid-Probe Viscotrabeculotomy in Neonatal-Onset Primary Congenital Glaucoma
Ahmed S. Elwehidy, Nader H.L. Bayoumi, Rasha M. Elzeini, Amr Abdelkader
Journal of Glaucoma. 2023; 32(9): 807
[Pubmed] | [DOI]
2 Refractive development I: Biometric changes during emmetropisation
Jos J. Rozema
Ophthalmic and Physiological Optics. 2023;
[Pubmed] | [DOI]
3 How many children with large clear cornea do not have glaucoma?
Nader Bayoumi, Eman Nabil Elsayed, Ahmed Saber
European Journal of Ophthalmology. 2023; : 1120672123
[Pubmed] | [DOI]
4 Risk of Developing Congenital Glaucoma in Premature Infants with Large Cup/Disc Ratio
Eun Jung Choi, Ji Seok Park, Ji Hye Shin, Mi-Young Jang, Young Hoon Hwang, Kyoung Nam Kim
Journal of the Korean Ophthalmological Society. 2023; 64(4): 323
[Pubmed] | [DOI]
5 Pediatric Glaucoma—From Screening, Early Detection to Management
Ruyue Shen, Venice S. W. Li, Mandy O. M. Wong, Poemen P. M. Chan
Children. 2023; 10(2): 181
[Pubmed] | [DOI]
6 Comparing the Use of Smartphone and Vernier Calipers for Corneal Diameter Measurement in Nigerian Term Neonates
N Mgbafulu, C Ezisi, H Ginger-Eke, R Umeh
Nigerian Journal of Clinical Practice. 2023; 26(11): 1758
[Pubmed] | [DOI]
7 Association Between Central Corneal Thickness and Growth Parameters Following Preterm
Gokhan CELIK, Murat GUNAY, Osman KIZILAY, Abdulhamit TÜTEN
Akdeniz Medical Journal. 2022; 8(1): 28
[Pubmed] | [DOI]
8 Normal range of ocular biometry in healthy children: A systemic review and meta-analysis of 33,559 individuals under seven years of age
Yanxian Chen, Wei Wang, Jun Wang, Xu Chen, Zhuoting Zhu, Jinying Li, Mingguang He
Ophthalmic and Physiological Optics. 2022;
[Pubmed] | [DOI]
9 Measurement and Analysis of Central Corneal Thickness at Different Postnatal Stages in Chinese Premature Infants
Chenyi Liu, Hanfei Wu, Jimeng Lao, Sulan Wu, Xiaoqiong Xu, Zhe Lv, Jianbo Mao
Journal of Ophthalmology. 2020; 2020: 1
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
Abstract
Methods
Results
Discussion
Conclusion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed2345    
    Printed72    
    Emailed0    
    PDF Downloaded244    
    Comments [Add]    
    Cited by others 9    

Recommend this journal